FDA Enforcement Class II Terminated

GE Centricity PACS Workstation, versions 3.1.1.x through 3.2.1.x Picture Archiving and Communication System Used as a primary diagnostic and analysis tool for diagnostic images by trained healthcare professionals, including radiologists, physicians, technologists, clinicians and nurses. Also used as a clinical review workstation throughout the healthcare facility. The workstation interface provides the user with a means to display, manipulate, archive, print and export images when connected with the Centricity PACS infrastructure.

Recall: Z-2579-2016 · Reported August 24, 2016

Enforcement

Recall Number
Z-2579-2016
Event ID
64785
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
GE Healthcare It
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 24, 2016
Initiation Date
November 15, 2012
Classification Date
August 17, 2016
Termination Date
December 13, 2017
Address
540 W Northwest Hwy, N/A, Barrington, IL, 60010-3051, United States

Description

GE Centricity PACS Workstation, versions 3.1.1.x through 3.2.1.x Picture Archiving and Communication System Used as a primary diagnostic and analysis tool for diagnostic images by trained healthcare professionals, including radiologists, physicians, technologists, clinicians and nurses. Also used as a clinical review workstation throughout the healthcare facility. The workstation interface provides the user with a means to display, manipulate, archive, print and export images when connected with the Centricity PACS infrastructure.

Reason

While merging exams in a test instance, merges carry forward in the production system, when the middle tier is configured to the wrong IP address of the Centricity Exam Manager. When a current and historical exam are opened at the same time, the system asynchronously tries to access the operation specifying how each image should be grouped, causing random image-series grouping errors.

Code Info

Centricity PACS versions 3.1.1.x through 3.2.1.x

Distribution

Worldwide distribution including US nationwide, Puerto Rico, Australia, Austria, Belgium, Bermuda, CANADA, Denmark, Egypt, ENGLAND, France, Germany, India, Ireland, Hong Kong, Korea, Kuwait, Italy, Malta, Netherlands, Portugal, Qatar, Saudi Arabia, SCOTLAND, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, Turkey, and UK.

Quantity

498 units