FDA Enforcement Class II Ongoing

ACL TOP Family 50 Series Models as follows: In vitro diagnostic clinical use in the hemostasis laboratory for coagulation and/or fibrinolysis testing in the assessment of thrombosis and/or hemostasis (1) ACL TOP 550 CTS Model Number: 00000280045; (2) ACL TOP 750 CTS Model Number: 00000280055; (3) ACL TOP 350 CTS Model Number: 00000280065; (4) ACL TOP 750 Model Number: 00000280015; (5) ACL TOP 750 LAS Model Number: 00000280035 (6) ACL TOP 970 CL(only available ROW)Model Number: 00000280097

Recall: Z-2578-2023 · Reported September 20, 2023

Enforcement

Recall Number
Z-2578-2023
Event ID
92831
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Instrumentation Laboratory
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
September 20, 2023
Initiation Date
July 13, 2023
Classification Date
September 12, 2023
Address
180 Hartwell Rd, Bedford, MA, 01730-2443, United States

Description

ACL TOP Family 50 Series Models as follows: In vitro diagnostic clinical use in the hemostasis laboratory for coagulation and/or fibrinolysis testing in the assessment of thrombosis and/or hemostasis (1) ACL TOP 550 CTS Model Number: 00000280045; (2) ACL TOP 750 CTS Model Number: 00000280055; (3) ACL TOP 350 CTS Model Number: 00000280065; (4) ACL TOP 750 Model Number: 00000280015; (5) ACL TOP 750 LAS Model Number: 00000280035 (6) ACL TOP 970 CL(only available ROW)Model Number: 00000280097

Reason

Sample misidentification could occur under specific conditions and patient management altered based on an incorrectly assigned result

Code Info

Model Names UDI (1)ACL TOP 550 CTS 08426950729242; (2)ACL TOP 750 CTS 08426950784067; (3)ACL TOP 350 CTS 08426950784081; (4) ACL TOP 750 08426950784074; (5) ACL TOP 750 LAS 08426950784050 (6) ACL TOP 970 CL 08430793045476 Software: 1)ACL TOP 550 CTS v6.3.0 or later; (2)ACL TOP 750 CTS v6.3.0 or later; (3)ACL TOP 350 CTS v6.3.0 or later; (4) ACL TOP 750 v6.3.0 or later; (5) ACL TOP 750 LAS v6.3.0 or later; (6) ACL TOP 970 CL Version 1.1.0 (OUS). UPDATE: 10/16/23 Affected Serial Number Range Start SN End SN ACL TOP 350 CTS 21113058 23074369; ACL TOP 550 CTS 21112046 23072623; ACL TOP 750 22010516 23060671; ACL TOP 750 CTS 21110860 23071015; ACL TOP 750 LAS 21120879 23071077

Distribution

Worldwide distribution - US Nationwide and the countries of Canada, Finland, Ireland, Lithuania, Malta, Qatar, Sweden.

Quantity

2,068 kits US: 6,270 units OUS