FDA Enforcement
Class I
Completed
SHERPA NX ACTIVE GUIDING CATHETER, 6F, EBU3.5, 90CM.070", REF SA6EBU35D. for cardiovascular use
Recall: Z-2577-2019
·
Reported October 9, 2019
Enforcement
- Recall Number
- Z-2577-2019
- Event ID
- 82525
- Classification
- Class I
- Status
- Completed
- Product Type
- Devices
- Firm
- Medtronic Vascular
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- October 9, 2019
- Initiation Date
- March 15, 2019
- Classification Date
- September 30, 2019
- Termination Date
- April 26, 2024
- Address
- 35 Cherry Hill Dr # 37a, N/A, Danvers, MA, 01923-2565, United States
Description
SHERPA NX ACTIVE GUIDING CATHETER, 6F, EBU3.5, 90CM.070", REF SA6EBU35D. for cardiovascular use
Reason
There is a potential for extensive loss of primary segment material exposing underlying stainless-steel braid wires on a subset of 6F Sherpa catheters.
Code Info
GTIN: 00613994823571 All Lot/Serial Numbers
Distribution
Worldwide distribution to US, Austria, Belgium, Bulgaria, Czech Republic, France, Germany, Greece, Hong Kong, Ireland, Italy, Kazakhstan, Luxembourg, Moldova, Netherlands, Poland, Portugal, Russian Federation, Saudi Arabia, Slovakia, South Africa, Spain, Sweden, Taiwan, and United Kingdom
Quantity
2 units