PERSONA The Personalized Knee System CR Tibial Articular Surface Provisional with the following: Left Size 3-9 CD Top; Left Size 3-11 EF Top; Left Size 7-12 GH Top; Left Size 9-12 J Top; Left Size 3-9 CD Top; Left Size 3-11 EF Top; Right Size 7-12 GH Top; and Right Size 9-12 J Top.
Enforcement
- Recall Number
- Z-2577-2014
- Event ID
- 68878
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Zimmer, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- September 10, 2014
- Initiation Date
- August 7, 2014
- Classification Date
- September 4, 2014
- Termination Date
- January 27, 2015
- Address
- 1800 W Center St, Warsaw, IN, 46580-2304, United States
Description
PERSONA The Personalized Knee System CR Tibial Articular Surface Provisional with the following: Left Size 3-9 CD Top; Left Size 3-11 EF Top; Left Size 7-12 GH Top; Left Size 9-12 J Top; Left Size 3-9 CD Top; Left Size 3-11 EF Top; Right Size 7-12 GH Top; and Right Size 9-12 J Top.
Reports of breakage of the Persona Knee Trial Articular Surface Provisionals (TASPs). The Persona primary knee surgical technique has been updated. This field action correction is to ensure that all distributors and users of this instrument are aware of the enhanced surgical techniques.
Part Numbers: 42517000410, 42517000510, 42517000610, 42517000710, 42527000410, 42527000510, 42527000610, and 42527000710. Lot numbers: 62024975, 62043546, 62144792, 62152725, 62165538, 62187318, 62337961, 62372407, 62386776, 62397159, 62405917, 62456263, 62499170, 62551386, 62565051, 62568272, 62572572, 62663168, 62691384, 62707494, 62717346, 62020920, 62144793, 62156313, 62165539, 62184228, 62353932, 62356176, 62385100, 62420242, 62468277, 62538412, 62564340, 62611088, 62612263, 62690056, 62702295, 62717337, 62736078, 62024976, 62054776, 62147232, 62177243, 62187319, 62187320, 62359057, 62364955, 62391222, 62414006, 62461233, 62558583, 62564345, 62608889, 62638657, 62646286, 62702749, 62724348, 62394279, 62443397, 62487490, 62492802, 62506001, 62563915, 62568271, 62704633, 62055686, 62144796, 62161537, 62162597, 62256286, 62351297, 62356578, 62385099, 62404282, 62424802, 62455897, 62538392, 62565037, 62595855, 62611083, 62663162, 62680121, 62686205, 62695286, 62724347, 62724744, 62019863, 62156317, 62165548, 62168386, 62184268, 62256288, 62344258, 62356586, 62391232, 62391233, 62455892, 62504877, 62551337, 62564335, 62597816, 62638685, 62646284, 62695878, 62036673, 62138318, 62149715, 62189326, 62189327, 62356593, 62377738, 62391236, 62420225, 62461232, 62564360, 62564361, 62595865, 62611086, 62680120, 62690951, 62710615, 62761109, 62394280, 62428338, 62444959, 62500931, and 62568278.
Worldwide Distribution: US (nationwide) including states of: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY; and countries of: AUSTRALIA, INDIA, JAPAN, KOREA, SINGAPORE, AUSTRIA, BELGIUM, FRANCE, GERMANY, ISRAEL, ITALY, LUXEMBOURG, NETHERLANDS, SAUDI ARABIA, SOUTH AFRICA, SPAIN, SWITZERLAND, UNITED KINGDOM, and UNITED ARAB EMIRATES.
77,737 total units