FDA Enforcement Class II Terminated

XLTEK EMU40EX EEG Headbox

Recall: Z-2570-2018 · Reported August 8, 2018

Enforcement

Recall Number
Z-2570-2018
Event ID
80239
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Natus Neurology DBA Excel Tech., Ltd. (XLTEK)
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 8, 2018
Initiation Date
May 15, 2018
Classification Date
July 28, 2018
Termination Date
May 6, 2024
Address
2568 Bristol Cir, N/A, Oakville, N/A, N/A, Canada

Description

XLTEK EMU40EX EEG Headbox

Reason

Possible degradation of an internal electronic component over time which may cause the XLTEK EMU 40EX Breakout box to fail. It may feel hot and plastic case may start melting.

Code Info

Part Numbers: 10406, 002933

Distribution

USA Distribution: AL, CA, CO, CT, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI MN, MO, NC, NE, NJ, NV, NY, OH, OR, PA, ,RI, SC, TN, TX, UT, VA, WA, WI. World wide Distribution: Canada, China, Cyprus, Germany, Hong Kong, Italy, Mexico, Saudi Arabia, Spain, United Kingdom

Quantity

462 Units Total