FDA Enforcement Class II Terminated

Fem IM Nail 14mmdx48cm right, Sterile, Rod, fixation, intramedullary and accessories- Specific Knee, Hip and Nail Implants

Recall: Z-2563-2017 · Reported June 21, 2017

Enforcement

Recall Number
Z-2563-2017
Event ID
76629
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Zimmer Biomet, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 21, 2017
Initiation Date
February 28, 2017
Classification Date
June 14, 2017
Termination Date
February 27, 2018
Address
1800 W Center St, N/A, Warsaw, IN, 46580-2304, United States

Description

Fem IM Nail 14mmdx48cm right, Sterile, Rod, fixation, intramedullary and accessories- Specific Knee, Hip and Nail Implants

Reason

Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.

Code Info

Item Number:00225748014, Lot Number Expiry Date Before: March 2027

Distribution

Worldwide Distribution - US (nationwide) Foreign: Argentina, Australia, Belgium, Brazil, Bridgend Wales, Canada, Chile, Colombia, Dominican Republic, England, Eschbach, France, India, Italy, Japan, Jordan, Korea, Malaysia, Mexico, Netherlands, Nicaragua, Panama, Peru, Russia, Saudi Arabia, Singapore, Spain, Switzerland, Taiwan, Thailand

Quantity

9867