Roche cobas IT 1000 Product Usage: Intended Use: An electronic device intended to store, retrieve, and process laboratory data.
Enforcement
- Recall Number
- Z-2563-2014
- Event ID
- 68912
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Roche Diagnostics Operations, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- September 10, 2014
- Initiation Date
- July 31, 2014
- Classification Date
- September 1, 2014
- Termination Date
- June 16, 2015
- Address
- 9115 Hague Rd, Indianapolis, IN, 46256-1025, United States
Description
Roche cobas IT 1000 Product Usage: Intended Use: An electronic device intended to store, retrieve, and process laboratory data.
Potential that test results would be assigned to the wrong Patient ID if cobas IT 1000 receives an ADT ADMIT, TRANSFER or MERGE event message that contains a genuine Patient ID (e.g. 999999), which is the same as what has been configured as the Generic Patient ID (999999 by default).
Part Numbers: 04420306001 Serial #'s: 2706, 2724, 2743, 2744, 2745, 2749, 2750, 2751, 2752, 2754, 2755, 2764, 2770, 2771, 2772, 2773, 2774, 2775, 2776, 2780, 2781, 2782, 2783, 2784, 2785, 2789, 2790, 2791, 2792, 2793, 2794, 2795, 2796, 2797, 2798, 2799, 2808, 2814, 2815, 2816, 2818, 2819, 2823, 2824, 2825, 2826, 2828, 2829, 2830, 2831, 2847, 2848, 2849, 2857, 2858, 2859, 2860, 2861, 2862, 2863, 2864, 2865, 2904, 2905, 2906, 2908, 2909, 2914, 2915, 2916, 2917, 2920, 2921, 2928, 2931, 2932, 2933, 2934, 2935, 2936, 2937, 2938, 2939, 2940, 2941, 2943, 2983, 2984, 2985, 2986, 2987, 2990, 2991, 2992, 2993, 2994, 2995, 2996, 2997, 2998, 3006, 3007, 3008, 3009, 3016, 3020, 3021, 3023, 3024, 3025, 3026, 3027, 3048, 3049, 3050, 3051, 3053, 3057, 3058, 3059, 3060, 3061, 3065, 3066, 3067, 3068, 3069, 3070, 3071, 3072, 3073, 3074, 3075, 3076, 3077, 3079, 3082, 3083, 3084, 3085, 3086, 3087, 3089, 3090, 3104, 3105, 3106, and 3107.
US Nationwide Distribution in the states of: AK, AL, AR, AZ, CA, CT, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MD, ME, MI, MO, NC, NE, NH, NJ, NY, OH, OR, PA, PR, SC, TN, TX, VA, VT, WA, WI, and WV.
148 Units