FDA Enforcement Class II Ongoing

Monaco RTP System, 5.40 Unity, radiation treatment planning system Product Usage: The Monaco system is used to make treatment plans for patients with prescriptions for external beam radiation therapy. The system calculates dose for photon, electron, and carbon treatment plans and displays, on-screen and in hard-copy, two- or three-dimensional radiation dose distributions inside patients for given treatment plan set-ups. The Monaco product line is intended for use in radiation treatment planning. It uses generally accepted methods for: -contouring -image manipulation -simulation -image fusion -plan optimization -QA and plan review

Recall: Z-2560-2019 · Reported October 2, 2019

Enforcement

Recall Number
Z-2560-2019
Event ID
83692
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Elekta Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
October 2, 2019
Initiation Date
September 4, 2019
Classification Date
September 25, 2019
Address
13723 Riverport Dr Ste 100, Maryland Heights, MO, 63043-4819, United States

Description

Monaco RTP System, 5.40 Unity, radiation treatment planning system Product Usage: The Monaco system is used to make treatment plans for patients with prescriptions for external beam radiation therapy. The system calculates dose for photon, electron, and carbon treatment plans and displays, on-screen and in hard-copy, two- or three-dimensional radiation dose distributions inside patients for given treatment plan set-ups. The Monaco product line is intended for use in radiation treatment planning. It uses generally accepted methods for: -contouring -image manipulation -simulation -image fusion -plan optimization -QA and plan review

Reason

It is possible that the forced electron density settings will be changed for some structures unintentionally and this can result in incorrect dose calculation.

Code Info

UDI (01)00858164002190(10) 5.40.00

Distribution

Worldwide Distribution - US Nationwide in the states of WI, OH, NY, OH, IN, IA, TX, and countries of Canada, China, Denmark, Germany, Greece, Hong Kong, Italy, Netherlands, Sweden, UK

Quantity

30 units