FDA Enforcement Class II Ongoing

Vercise Deep Brain Stimulation Systems Surgical Implant Manual, Document Number: 92328632-09

Recall: Z-2558-2025 · Reported September 24, 2025

Enforcement

Recall Number
Z-2558-2025
Event ID
97144
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Boston Scientific Neuromodulation Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
September 24, 2025
Initiation Date
July 8, 2025
Classification Date
September 12, 2025
Address
25155 Rye Canyon Loop, N/A, Valencia, CA, 91355-5004, United States

Description

Vercise Deep Brain Stimulation Systems Surgical Implant Manual, Document Number: 92328632-09

Reason

Wire break(s) have occurred in rechargeable deep brain stimulation implantable pulse generators, implanted in the submuscular pectoral location. Wire breaks prevent successful stimulation, thus requiring device removal/replacement and users may experience high monopolar impedances, undesired sensation, loss of therapy, return of symptoms and/or Bluetooth connectivity challenges.

Code Info

UDI-DI: 08714729985044, 08714729985051. Document Number/Revision: 92328632-09, Rev A

Distribution

Worldwide - US Nationwide distribution including in the states of MN, FL, CO, MD, NY, NE, PA, WI, AZ, TN, MO, TX, MI, MT, MA, WV, AL, VA, NC, CA, OR, WA, LA, OH, NJ, NH, SD, GA, CT, NV, UT, IN, ID, IL, ME, SC, DC, OK, KY, AK, HI, RI, AR, KS, NM and the countries of Argentina, Brazil, Chile, Ecuador, Colombia , Mexico, Canada, Japan, Algeria, Austria, Azerbaijan, Belgium, Bulgaria, Croatia, Denmark, Egypt, Finland, France, Georgia, Germany, Greece, Hungary, Iceland, Iran, Ireland, Israel, Italy, Jordan, Kazakhstan, Kuwait, Lebanon, Libya, Luxembourg, Malta, Morocco, Netherlands, Norway, Pakistan, Poland, Portugal, Romania, Russian Fed., Saudi Arabia, Serbia, Slovakia, South Africa, Spain, Sweden, Switzerland, Turkiye, UAE, Ukraine, United Kingdom, Hong Kong, China, Singapore, Korea, Taiwan , India, Thailand.

Quantity

25,260