Vercise Deep Brain Stimulation Systems Surgical Implant Manual, Document Number: 92328632-09
Enforcement
- Recall Number
- Z-2558-2025
- Event ID
- 97144
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Boston Scientific Neuromodulation Corporation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- September 24, 2025
- Initiation Date
- July 8, 2025
- Classification Date
- September 12, 2025
- Address
- 25155 Rye Canyon Loop, N/A, Valencia, CA, 91355-5004, United States
Description
Vercise Deep Brain Stimulation Systems Surgical Implant Manual, Document Number: 92328632-09
Wire break(s) have occurred in rechargeable deep brain stimulation implantable pulse generators, implanted in the submuscular pectoral location. Wire breaks prevent successful stimulation, thus requiring device removal/replacement and users may experience high monopolar impedances, undesired sensation, loss of therapy, return of symptoms and/or Bluetooth connectivity challenges.
UDI-DI: 08714729985044, 08714729985051. Document Number/Revision: 92328632-09, Rev A
Worldwide - US Nationwide distribution including in the states of MN, FL, CO, MD, NY, NE, PA, WI, AZ, TN, MO, TX, MI, MT, MA, WV, AL, VA, NC, CA, OR, WA, LA, OH, NJ, NH, SD, GA, CT, NV, UT, IN, ID, IL, ME, SC, DC, OK, KY, AK, HI, RI, AR, KS, NM and the countries of Argentina, Brazil, Chile, Ecuador, Colombia , Mexico, Canada, Japan, Algeria, Austria, Azerbaijan, Belgium, Bulgaria, Croatia, Denmark, Egypt, Finland, France, Georgia, Germany, Greece, Hungary, Iceland, Iran, Ireland, Israel, Italy, Jordan, Kazakhstan, Kuwait, Lebanon, Libya, Luxembourg, Malta, Morocco, Netherlands, Norway, Pakistan, Poland, Portugal, Romania, Russian Fed., Saudi Arabia, Serbia, Slovakia, South Africa, Spain, Sweden, Switzerland, Turkiye, UAE, Ukraine, United Kingdom, Hong Kong, China, Singapore, Korea, Taiwan , India, Thailand.
25,260