FDA Enforcement Class II Terminated

Cordis EMPIRA NC RX PTCA Dilatation Catheter

Recall: Z-2556-2014 · Reported September 3, 2014

Enforcement

Recall Number
Z-2556-2014
Event ID
68834
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Cordis Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
September 3, 2014
Initiation Date
June 23, 2014
Classification Date
August 28, 2014
Termination Date
December 14, 2016
Address
14201 Nw 60th Ave, N/A, Miami Lakes, FL, 33014-2802, United States

Description

Cordis EMPIRA NC RX PTCA Dilatation Catheter

Reason

The balloon may exhibit resistance to the removal of the balloon cover. Weakening of the affected product in the region of the proximal balloon seal may occur when excessive force is applied to remove the balloon cover. This may lead to balloon burst, shaft separation or deflation difficulties, which could result in procedural delay, vessel trauma or surgery.

Code Info

63 Catalog numbers: 75R06200N, 75R06225N, 75R06250N, 75R06275N, 75R06300N, 75R06325N, 75R06350N, 75R06375N, 75R06400N, 75R10200N, 75R10225N, 75R10250N, 75R10275N, 75R10300N, 75R10325N, 75R10350N, 75R10375N, 75R10400N, 75R12200N, 75R12225N, 75R12250N, 75R12275N, 75R12300N, 75R12325N, 75R12350N, 75R12375N, 75R12400N, 75R15200N, 75R15225N, 75R15250N, 75R15275N, 75R15300N, 75R15325N, 75R15350N, 75R15375N, 75R15400N, 75R20200N, 75R20225N, 75R20250N, 75R20275N, 75R20300N, 75R20325N, 75R20350N, 75R20375N, 75R20400N, 75R25200N, 75R25225N, 75R25250N, 75R25275N, 75R25300N, 75R25325N, 75R25350N, 75R25375N, 75R25400N, 75R25200N, 75R25225N, 75R25250N, 75R25275N, 75R25300N, 75R25325N, 75R25350N, 75R25375N, and 75R25400N.

Distribution

Worldwide Distribution - USA including Puerto Rico and the states of AR, AZ, CA, CO, CT, DC, FL, GA, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WV, WY and Hawaii., and the countries of Argentina, Brazil, Canada, Chile, Colombia, Mexico, Panama and Venezuela.

Quantity

250,000 units