FDA Enforcement Class II Terminated

The Thermedx Fluid Management System is designed to provide irrigation fluids to irrigate, distend, and debride while warming the fluid to assist in the prevention of peri-operative hypothermia. The fluid is warmed via an internal infrared mechanism. Additionally the system will hold suction canisters for the collection of used irrigant.

Recall: Z-2555-2014 · Reported September 3, 2014

Enforcement

Recall Number
Z-2555-2014
Event ID
68986
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Thermedx LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
September 3, 2014
Initiation Date
February 10, 2014
Classification Date
August 27, 2014
Termination Date
August 4, 2015
Address
31200 Solon Rd, Unit 1, Solon, OH, 44139-3556, United States

Description

The Thermedx Fluid Management System is designed to provide irrigation fluids to irrigate, distend, and debride while warming the fluid to assist in the prevention of peri-operative hypothermia. The fluid is warmed via an internal infrared mechanism. Additionally the system will hold suction canisters for the collection of used irrigant.

Reason

A vendor changed the manufacturing process of a component used in the canister ring that may affect the performance of the ring. The deficit display may be higher than the actual deficit when the fluid collection canister is fully loaded.

Code Info

Model P4000 Serial numbers affected: 20110004, 20110007, 20110015, 20110025, 20130001-20130032, 20140001-20140005

Distribution

Nationwide Distribution including HI, MS, NY, OH, PA, VA, WA, WV, and IL.

Quantity

21