FDA Enforcement Class II Terminated

CURBELL MEDICAL CareSense Wireless Advanced LCD Bed and Chair Monitor, MODEL # CSM-BC400; --- COMMON/USUAL NAME: Bed Patient Monitor; CLASSIFICATION NAME: Bed Patient Monitor --- Monitor is connected to a sensor pad that monitors the patient if they attempt to arise from a bed, chair, floor mat, or toilet seat. The monitor will sound an alarm.

Recall: Z-2554-2014 · Reported September 3, 2014

Enforcement

Recall Number
Z-2554-2014
Event ID
68961
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Curbell Medical, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
September 3, 2014
Initiation Date
August 5, 2014
Classification Date
August 26, 2014
Termination Date
February 28, 2017
Address
20 Centre Dr, N/A, Orchard Park, NY, 14127-4102, United States

Description

CURBELL MEDICAL CareSense Wireless Advanced LCD Bed and Chair Monitor, MODEL # CSM-BC400; --- COMMON/USUAL NAME: Bed Patient Monitor; CLASSIFICATION NAME: Bed Patient Monitor --- Monitor is connected to a sensor pad that monitors the patient if they attempt to arise from a bed, chair, floor mat, or toilet seat. The monitor will sound an alarm.

Reason

The firm became aware of a potential problem that was initiated by a customer complaint. After consultation with the manufacturer, it was discovered that a resistor was incorrectly placed within the circuit board on the monitor. This change to the resistor was a planned change to address a product improvement (improve battery drain). However, this modification rendered the monitor's alarm via the nurse call communication port to not function.

Code Info

Lot 0414

Distribution

Nationwide, Canada, Ireland, Italy

Quantity