FDA Enforcement Class II Terminated

Siemens Dimension EXL Integrated Chemistry System in vitro diagnostic devices intended to measure a variety of analytes in human body fluids. All Serial #s of material #s: 10636929, 10636928, 10486890, 10472176, 10639337, and 10639338.

Recall: Z-2553-2014 · Reported September 3, 2014

Enforcement

Recall Number
Z-2553-2014
Event ID
68854
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Siemens Healthcare Diagnostics, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
September 3, 2014
Initiation Date
July 16, 2014
Classification Date
August 26, 2014
Termination Date
December 16, 2014
Address
500 Gbc Dr Ms 514, PO BOX 6101, Newark, DE, 19702-2466, United States

Description

Siemens Dimension EXL Integrated Chemistry System in vitro diagnostic devices intended to measure a variety of analytes in human body fluids. All Serial #s of material #s: 10636929, 10636928, 10486890, 10472176, 10639337, and 10639338.

Reason

The Reagent Lid hinge may lose its effectiveness and slowly shift downward during maintenance procedures, potentially resulting in physical contact and injury to the operator.

Code Info

All Serial #s of material #s: 10636929, 10636928, 10486890, 10472176, 10639337, and 10639338

Distribution

Nationwide and internationally, including Canada and Mexico

Quantity

2757