Siemens Dimension EXL Integrated Chemistry System in vitro diagnostic devices intended to measure a variety of analytes in human body fluids. All Serial #s of material #s: 10636929, 10636928, 10486890, 10472176, 10639337, and 10639338.
Recall: Z-2553-2014
·
Reported September 3, 2014
Enforcement
- Recall Number
- Z-2553-2014
- Event ID
- 68854
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Siemens Healthcare Diagnostics, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- September 3, 2014
- Initiation Date
- July 16, 2014
- Classification Date
- August 26, 2014
- Termination Date
- December 16, 2014
- Address
- 500 Gbc Dr Ms 514, PO BOX 6101, Newark, DE, 19702-2466, United States
Description
Siemens Dimension EXL Integrated Chemistry System in vitro diagnostic devices intended to measure a variety of analytes in human body fluids. All Serial #s of material #s: 10636929, 10636928, 10486890, 10472176, 10639337, and 10639338.
Reason
The Reagent Lid hinge may lose its effectiveness and slowly shift downward during maintenance procedures, potentially resulting in physical contact and injury to the operator.
Code Info
All Serial #s of material #s: 10636929, 10636928, 10486890, 10472176, 10639337, and 10639338
Distribution
Nationwide and internationally, including Canada and Mexico
Quantity
2757