FDA Enforcement Class II Completed

QC (Cs-137) screened nanoDot D2DNS, Model Number 03067-OTO; radiation monitoring dosimeter used with the microSTAR readers

Recall: Z-2552-2023 · Reported September 20, 2023

Enforcement

Recall Number
Z-2552-2023
Event ID
92647
Classification
Class II
Status
Completed
Product Type
Devices
Firm
Landauer
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
September 20, 2023
Initiation Date
July 12, 2023
Classification Date
September 12, 2023
Address
2 Science Rd, Glenwood, IL, 60425-1531, United States

Description

QC (Cs-137) screened nanoDot D2DNS, Model Number 03067-OTO; radiation monitoring dosimeter used with the microSTAR readers

Reason

LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.

Code Info

UDI/DI 0860003399903, all batch numbers

Distribution

Worldwide

Quantity

0