FDA Enforcement Class II Terminated

Trauma Large Screws: CANN BONE SCREW 7.0MMX100MMLG CANN BONE SCREW 7.0X105LG CANN BONE SCREW 7.0X105LG CANN BONE SCREW 7.0MMX105MMLG CANN BONE SCREW 7.0X110LG CANN BONE SCREW 7.0X110LG CANN BONE SCREW 7.0MMX110MMLG CANN BONE SCREW 7.0X115LG CANN BONE SCREW 7.0X115LG CANN BONE SCREW 7.0MMX115MMLG CANN BONE SCREW 7.0X120LG CANN BONE SCREW 7.0X120LG CANN BONE SCREW 7.0MMX120MMLG CANN BONE SCREW 7.0X125LG CANN BONE SCREW 7.0MMX125MMLG CANN BONE SCREW 7.0X130LG CANN BONE SCREW 7.0MMX130MMLG RECON SCR LGE 5.5MMDX105MM RECON SCR LGE 5.5MMDX110MM RECON SCR LGE 5.5MMDX115MM RECON SCR LGE 5.5MMDX120MM RECON SCR LGE 5.5MMDX125MM RECON SCR LGE 5.5MMDX130MM ITST ANTI-ROT SCR 6.5MMDX105MM ITST ANTI-ROT SCR 6.5MMDX110MM PERI. SCR 3.5MM X105MML PERI SCR. 6.5MM X 100MM L PERI SCR. 6.5MM X 140MM L PERI SCR. 6.5MM X 100MM L PERI SCR. 6.5MM X 100MM L 3.5MM CORT SCR X 105MM SELFTAP 3.5MM CORT SCR X 110MM SELFTAP 5.5MM X 100MM CANN LKNG SCREW Z NAIL 6.0 X 105 CANC SCREW PT Z NAIL 6.0 X 110 CANC SCREW PT Z NAIL 6.0 X 115 CANC SCREW PT Z NAIL 6.0 X 120 CANC SCREW PT Z NAIL 6.0 X 125 CANC SCREW PT Z NAIL 6.0 X 130 CANC SCREW PT Z NAIL 6.0 X 135 CANC SCREW PT Z NAIL 6.0 X 140 CANC SCREW PT Z NAIL 6.0 X 105 CANC SCREW FA Z NAIL 6.0 X 110 CANC SCREW FA Z NAIL 6.0 X 115 CANC SCREW FA Z NAIL 6.0 X 120 CANC SCREW FA Z NAIL 6.0 X 125 CANC SCREW FA Z NAIL 6.0 X 130 CANC SCREW FA Z NAIL 6.0 X 135 CANC SCREW FA Z NAIL 6.0 X 140 CANC SCREW FA Z NAIL 6.0 X 105 CANC SCREW PT Z NAIL 6.0 X 110 CANC SCREW PT Z NAIL 6.0 X 115 CANC SCREW PT Z NAIL 6.0 X 120 CANC SCREW PT Z NAIL 6.0 X 125 CANC SCREW PT Z NAIL 6.0 X 130 CANC SCREW PT Z NAIL 6.0 X 135 CANC SCREW PT Z NAIL 6.0 X 140 CANC SCREW PT Z NAIL 6.0 X 105 CANC SCREW FA Z NAIL 6.0 X 110 CANC SCREW FA Z NAIL 6.0 X 115CANC SCREW FA Z NAIL 6.0 X 120CANC SCREW FA Z NAIL 6.0 X 125 CANC SCREW FA Z NAIL 6.0 X 130 CANC SCREW FA Z NAIL 6.0 X 135 CANC SCREW FA Z NAIL 6.0 X 140 CANC SCREW FA 3.5X105 CORT SCR 3.5 X 105 CORT SCR SELF TAP 3.5X110 CORT SCR 3.5 X 110 CORT SCR SELF TAP 3.5X115 CORT SCR 3.5 X 115 CORT SCR SELF TAP 3.5X120 CORT SCR 3.5 X 120 CORT SCR SELF TAP 3.5X125 CORT SCR 3.5X130 CORT SCR 3.5X135 CORT SCR 3.5X140 CORT SCR 3.5X145 CORT SCR 3.5X150 CORT SCR 4.5 X 105 CORT SCR SELF TAP 4.5 X 110 CORT SCR SELF TAP 4.5 X 120 CORT SCR SELF TAP 6.5X105 CANC SCR, FULLY THDED 6.5X105 CANC SCR, 16MM THD 6.5X105 CANC SCR, 32MM THD 6.5X110 CANC SCR, FULLY THDED 6.5X110 CANC SCR, 16MM THD 6.5X110 CANC SCR, 32MM THD 6.5X115 CANC SCR, 32MM THD 6.5X120 CANC SCR, 32MM THD

Recall: Z-2549-2014 · Reported September 3, 2014

Enforcement

Recall Number
Z-2549-2014
Event ID
68818
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Zimmer, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
September 3, 2014
Initiation Date
July 22, 2014
Classification Date
August 26, 2014
Termination Date
September 15, 2015
Address
1800 W Center St, N/A, Warsaw, IN, 46580-2304, United States

Description

Trauma Large Screws: CANN BONE SCREW 7.0MMX100MMLG CANN BONE SCREW 7.0X105LG CANN BONE SCREW 7.0X105LG CANN BONE SCREW 7.0MMX105MMLG CANN BONE SCREW 7.0X110LG CANN BONE SCREW 7.0X110LG CANN BONE SCREW 7.0MMX110MMLG CANN BONE SCREW 7.0X115LG CANN BONE SCREW 7.0X115LG CANN BONE SCREW 7.0MMX115MMLG CANN BONE SCREW 7.0X120LG CANN BONE SCREW 7.0X120LG CANN BONE SCREW 7.0MMX120MMLG CANN BONE SCREW 7.0X125LG CANN BONE SCREW 7.0MMX125MMLG CANN BONE SCREW 7.0X130LG CANN BONE SCREW 7.0MMX130MMLG RECON SCR LGE 5.5MMDX105MM RECON SCR LGE 5.5MMDX110MM RECON SCR LGE 5.5MMDX115MM RECON SCR LGE 5.5MMDX120MM RECON SCR LGE 5.5MMDX125MM RECON SCR LGE 5.5MMDX130MM ITST ANTI-ROT SCR 6.5MMDX105MM ITST ANTI-ROT SCR 6.5MMDX110MM PERI. SCR 3.5MM X105MML PERI SCR. 6.5MM X 100MM L PERI SCR. 6.5MM X 140MM L PERI SCR. 6.5MM X 100MM L PERI SCR. 6.5MM X 100MM L 3.5MM CORT SCR X 105MM SELFTAP 3.5MM CORT SCR X 110MM SELFTAP 5.5MM X 100MM CANN LKNG SCREW Z NAIL 6.0 X 105 CANC SCREW PT Z NAIL 6.0 X 110 CANC SCREW PT Z NAIL 6.0 X 115 CANC SCREW PT Z NAIL 6.0 X 120 CANC SCREW PT Z NAIL 6.0 X 125 CANC SCREW PT Z NAIL 6.0 X 130 CANC SCREW PT Z NAIL 6.0 X 135 CANC SCREW PT Z NAIL 6.0 X 140 CANC SCREW PT Z NAIL 6.0 X 105 CANC SCREW FA Z NAIL 6.0 X 110 CANC SCREW FA Z NAIL 6.0 X 115 CANC SCREW FA Z NAIL 6.0 X 120 CANC SCREW FA Z NAIL 6.0 X 125 CANC SCREW FA Z NAIL 6.0 X 130 CANC SCREW FA Z NAIL 6.0 X 135 CANC SCREW FA Z NAIL 6.0 X 140 CANC SCREW FA Z NAIL 6.0 X 105 CANC SCREW PT Z NAIL 6.0 X 110 CANC SCREW PT Z NAIL 6.0 X 115 CANC SCREW PT Z NAIL 6.0 X 120 CANC SCREW PT Z NAIL 6.0 X 125 CANC SCREW PT Z NAIL 6.0 X 130 CANC SCREW PT Z NAIL 6.0 X 135 CANC SCREW PT Z NAIL 6.0 X 140 CANC SCREW PT Z NAIL 6.0 X 105 CANC SCREW FA Z NAIL 6.0 X 110 CANC SCREW FA Z NAIL 6.0 X 115CANC SCREW FA Z NAIL 6.0 X 120CANC SCREW FA Z NAIL 6.0 X 125 CANC SCREW FA Z NAIL 6.0 X 130 CANC SCREW FA Z NAIL 6.0 X 135 CANC SCREW FA Z NAIL 6.0 X 140 CANC SCREW FA 3.5X105 CORT SCR 3.5 X 105 CORT SCR SELF TAP 3.5X110 CORT SCR 3.5 X 110 CORT SCR SELF TAP 3.5X115 CORT SCR 3.5 X 115 CORT SCR SELF TAP 3.5X120 CORT SCR 3.5 X 120 CORT SCR SELF TAP 3.5X125 CORT SCR 3.5X130 CORT SCR 3.5X135 CORT SCR 3.5X140 CORT SCR 3.5X145 CORT SCR 3.5X150 CORT SCR 4.5 X 105 CORT SCR SELF TAP 4.5 X 110 CORT SCR SELF TAP 4.5 X 120 CORT SCR SELF TAP 6.5X105 CANC SCR, FULLY THDED 6.5X105 CANC SCR, 16MM THD 6.5X105 CANC SCR, 32MM THD 6.5X110 CANC SCR, FULLY THDED 6.5X110 CANC SCR, 16MM THD 6.5X110 CANC SCR, 32MM THD 6.5X115 CANC SCR, 32MM THD 6.5X120 CANC SCR, 32MM THD

Reason

Zimmer is recalling sterile-packaged implants and instruments packaged in six different package configurations due to packaging design verification test failures. Specifically, multiple test samples from each of the six configurations failed simulated distribution and shipping testing when the devices either breached inner or outer package trays or seal integrity was lost.

Code Info

Item Numbers 00114610099, 00114610500, 00114610532, 00114610599, 00114611000, 00114611032, 00114611099, 00114611500, 00114611532, 00114611599, 00114612000, 00114612032, 00114612099, 00114612500, 00114612599, 00114613000, 00114613099, 00225210555, 00225211055, 00225211555, 00225212055, 00225212555, 00225213055, 00225710565, 00225711065, 47234712005, 47234712400, 47234712440, 47234712500, 47234712600, 47234810535, 47234811035, 47235910055, 47248310560, 47248311060, 47248311560, 47248312060, 47248312560, 47248313060, 47248313560, 47248314060, 47248410560, 47248411060, 47248411560, 47248412060, 47248412560, 47248413060, 47248413560, 47248414060, 47348310560, 47348311060, 47348311560, 47348312060, 47348312560, 47348313060, 47348313560, 47348314060, 47348410560, 47348411060, 47348411560, 47348412060, 47348412560, 47348413060, 47348413560, 47348414060, 47483510500, 47483510501, 47483511000, 47483511001, 47483511500, 47483511501, 47483512000, 47483512001, 47483512500, 47483513000, 47483513500, 47483514000, 47483514500, 47483515000, 47484510501, 47484511001, 47484512001, 47486510500, 47486510501, 47486510502, 47486511000, 47486511001, 47486511002, 47486511502, 47486512002 EXPIRATION DATE PRIOR TO 2024-07

Distribution

Worldwide Distribution.

Quantity

972,125 units