FDA Enforcement
Class II
Completed
nanoDot D2DXS, Model Numbers: a) 03063-1MO; b) 03063-3MO; c) 03063-OTO; d) 03063-SMO; radiation monitoring dosimeter used with the microSTAR readers
Recall: Z-2548-2023
·
Reported September 20, 2023
Enforcement
- Recall Number
- Z-2548-2023
- Event ID
- 92647
- Classification
- Class II
- Status
- Completed
- Product Type
- Devices
- Firm
- Landauer
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- September 20, 2023
- Initiation Date
- July 12, 2023
- Classification Date
- September 12, 2023
- Address
- 2 Science Rd, Glenwood, IL, 60425-1531, United States
Description
nanoDot D2DXS, Model Numbers: a) 03063-1MO; b) 03063-3MO; c) 03063-OTO; d) 03063-SMO; radiation monitoring dosimeter used with the microSTAR readers
Reason
LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.
Code Info
UDI/DI 0860003399903, all batch numbers
Distribution
Worldwide
Quantity
68007 units