FDA Enforcement Class II Terminated

Endo & Unipolar Femoral Heads: ENDO FEMORAL HEAD 41MM DIA ENDO FEMORAL HEAD 42MM DIA ENDO FEMORAL HEAD 43MM DIA ENDO FEMORAL HEAD 44MM DIA ENDO FEMORAL HEAD 45MM DIA ENDO FEMORAL HEAD 46MM DIA ENDO FEMORAL HEAD 47MM DIA ENDO FEMORAL HEAD 48MM DIA ENDO FEMORAL HEAD 49MM DIA ENDO FEMORAL HEAD 50MM DIA ENDO FEMORAL HEAD 51MM DIA ENDO FEMORAL HEAD 52MM DIA ENDO FEMORAL HEAD 53MM DIA ENDO FEMORAL HEAD 54MM DIA ENDO FEMORAL HEAD 55MM DIA ENDO FEMORAL HEAD 57MM DIA ENDO FEMORAL HEAD 60MM DIA ENDO FEMORAL HEAD 63MM DIA ENDO FEM HEAD 41MM ENDO FEM HEAD 42MM ENDO FEM HEAD 43MM ENDO PROS HEAD 44MM ENDO FEM HEAD 45MM ENDO FEM HEAD 46MM ENDO FEM HEAD 47MM ENDO FEM HEAD 48MM ENDO FEM HEAD 49MM ENDO FEM HEAD 50MM ENDO FEM HEAD 51MM ENDO FEM HEAD 52MM ENDO FEM HEAD 53MM ENDO FEM HEAD 54MM ENDO FEM HEAD 55MM ENDO FEM HEAD 57MM ENDO FEM HEAD 60MM ENDO FEM HEAD 63MM UNIPOLAR COMPONENT, 41MM UNIPOLAR COMPONENT, 42MM UNIPOLAR COMPONENT, 43MM UNIPOLAR COMPONENT, 44MM UNIPOLAR COMPONENT, 45MM UNIPOLAR COMPONENT, 46MM UNIPOLAR COMPONENT, 47MM UNIPOLAR COMPONENT, 48MM UNIPOLAR COMPONENT, 49MM UNIPOLAR COMPONENT, 50MM UNIPOLAR COMPONENT, 51MM UNIPOLAR COMPONENT, 52MM UNIPOLAR COMPONENT, 53MM UNIPOLAR COMPONENT, 54MM UNIPOLAR COMPONENT, 55MM UNIPOLAR COMPONENT, 56MM UNIPOLAR COMPONENT, 57MM UNIPOLAR COMPONENT, 58MM UNIPOLAR COMPONENT, 59MM UNIPOLAR COMPONENT, 60MM

Recall: Z-2547-2014 · Reported September 3, 2014

Enforcement

Recall Number
Z-2547-2014
Event ID
68818
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Zimmer, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
September 3, 2014
Initiation Date
July 22, 2014
Classification Date
August 26, 2014
Termination Date
September 15, 2015
Address
1800 W Center St, N/A, Warsaw, IN, 46580-2304, United States

Description

Endo & Unipolar Femoral Heads: ENDO FEMORAL HEAD 41MM DIA ENDO FEMORAL HEAD 42MM DIA ENDO FEMORAL HEAD 43MM DIA ENDO FEMORAL HEAD 44MM DIA ENDO FEMORAL HEAD 45MM DIA ENDO FEMORAL HEAD 46MM DIA ENDO FEMORAL HEAD 47MM DIA ENDO FEMORAL HEAD 48MM DIA ENDO FEMORAL HEAD 49MM DIA ENDO FEMORAL HEAD 50MM DIA ENDO FEMORAL HEAD 51MM DIA ENDO FEMORAL HEAD 52MM DIA ENDO FEMORAL HEAD 53MM DIA ENDO FEMORAL HEAD 54MM DIA ENDO FEMORAL HEAD 55MM DIA ENDO FEMORAL HEAD 57MM DIA ENDO FEMORAL HEAD 60MM DIA ENDO FEMORAL HEAD 63MM DIA ENDO FEM HEAD 41MM ENDO FEM HEAD 42MM ENDO FEM HEAD 43MM ENDO PROS HEAD 44MM ENDO FEM HEAD 45MM ENDO FEM HEAD 46MM ENDO FEM HEAD 47MM ENDO FEM HEAD 48MM ENDO FEM HEAD 49MM ENDO FEM HEAD 50MM ENDO FEM HEAD 51MM ENDO FEM HEAD 52MM ENDO FEM HEAD 53MM ENDO FEM HEAD 54MM ENDO FEM HEAD 55MM ENDO FEM HEAD 57MM ENDO FEM HEAD 60MM ENDO FEM HEAD 63MM UNIPOLAR COMPONENT, 41MM UNIPOLAR COMPONENT, 42MM UNIPOLAR COMPONENT, 43MM UNIPOLAR COMPONENT, 44MM UNIPOLAR COMPONENT, 45MM UNIPOLAR COMPONENT, 46MM UNIPOLAR COMPONENT, 47MM UNIPOLAR COMPONENT, 48MM UNIPOLAR COMPONENT, 49MM UNIPOLAR COMPONENT, 50MM UNIPOLAR COMPONENT, 51MM UNIPOLAR COMPONENT, 52MM UNIPOLAR COMPONENT, 53MM UNIPOLAR COMPONENT, 54MM UNIPOLAR COMPONENT, 55MM UNIPOLAR COMPONENT, 56MM UNIPOLAR COMPONENT, 57MM UNIPOLAR COMPONENT, 58MM UNIPOLAR COMPONENT, 59MM UNIPOLAR COMPONENT, 60MM

Reason

Zimmer is recalling sterile-packaged implants and instruments packaged in six different package configurations due to packaging design verification test failures. Specifically, multiple test samples from each of the six configurations failed simulated distribution and shipping testing when the devices either breached inner or outer package trays or seal integrity was lost.

Code Info

Item Numbers 00781804100, 00781804200, 00781804300, 00781804400, 00781804500, 00781804600, 00781804700, 00781804800, 00781804900, 00781805000, 00781805100, 00781805200, 00781805300, 00781805400, 00781805500, 00781805700, 00781806000, 00781806300, 00902604101, 00902604201, 00902604301, 00902604401, 00902604501, 00902604601, 00902604701, 00902604801, 00902604901, 00902605001, 00902605101, 00902605201, 00902605301, 00902605401, 00902605501, 00902605701, 00902606001, 00902606301, 451000041, 451000042, 451000043, 451000044, 451000045, 451000046, 451000047, 451000048, 451000049, 451000050, 451000051, 451000052, 451000053, 451000054, 451000055, 451000056, 451000057, 451000058, 451000059, 451000060 EXPIRATION DATE PRIOR TO 2024-01

Distribution

Worldwide Distribution.

Quantity

972,125 units