FDA Enforcement Class II Terminated

Zimmer Gender Solution Patello-Femoral Joint (PFJ) System Patello-Femoral Trochlea Component, Precoat, Size 4, Left. Knee prosthesis.

Recall: Z-2545-2014 · Reported September 3, 2014

Enforcement

Recall Number
Z-2545-2014
Event ID
68934
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Zimmer, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
September 3, 2014
Initiation Date
July 30, 2014
Classification Date
August 25, 2014
Termination Date
January 20, 2015
Address
1800 W Center St, N/A, Warsaw, IN, 46580-2304, United States

Description

Zimmer Gender Solution Patello-Femoral Joint (PFJ) System Patello-Femoral Trochlea Component, Precoat, Size 4, Left. Knee prosthesis.

Reason

Internal review found that five (5) PFJ implant forgings were identified as have forged laps during receiving inspection process for raw material. These five forgings were not segregated and were subsequently issued to final manufacturing under Part 00-5926-014-01; lot 62455332. A forged lap is a defect that forms when metal folds over itself during forging.

Code Info

00-5926-014-01; Lot 62455332

Distribution

Worldwide Distribution - US (nationwide) in the states of HI, CA, GA, MI, and IL and the countries of TAIWAN, BELGIUM, AUSTRALIA, FRANCE, JAPAN, and UK.

Quantity

15 units