Subcutaneous Implantable Cardioverter Defibrillator (S-ICD) System, labeled as EMBLEM MRI S-ICD Pulse Generator
Enforcement
- Recall Number
- Z-2543-2019
- Event ID
- 83558
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Boston Scientific Corporation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- September 25, 2019
- Initiation Date
- August 21, 2019
- Classification Date
- September 13, 2019
- Address
- 4100 Hamline Ave N Bldg 3, N/A, Saint Paul, MN, 55112-5700, United States
Description
Subcutaneous Implantable Cardioverter Defibrillator (S-ICD) System, labeled as EMBLEM MRI S-ICD Pulse Generator
Potential need for device replacement - Elective Replacement Indicator (ERI) or End of Life (EOL) - earlier than expected due to compromised performance of an electrical component causing accelerated battery depletion.
Model A219, Serial Number (Use Before Date) - 213636 (7/21/2019), 213637 (7/21/2019), 213638 (7/21/2019), 213639 (7/21/2019), 213640 (7/21/2019), 213641 (7/21/2019), 213642 (7/21/2019), 213643 (7/21/2019), 213644 (7/21/2019), 213645 (7/21/2019), 213646 (7/21/2019), 213647 (7/21/2019), 213648 (7/21/2019), 213649 (7/21/2019), 213650 (7/21/2019), 213651 (7/21/2019), 213652 (7/21/2019), 213653 (7/21/2019), 213654 (7/21/2019), 213655 (7/21/2019), 213656 (7/21/2019), 213657 (7/21/2019), 213658 (7/21/2019), 213659 (7/21/2019), 213660 (7/21/2019), 213661 (7/21/2019), 213662 (7/21/2019), 213663 (7/21/2019), 213665 (7/21/2019), 213666 (7/21/2019), 213667 (7/21/2019), 213668 (7/21/2019), 213669 (7/21/2019), 213670 (7/21/2019), 213671 (7/21/2019), 213672 (7/21/2019), 213673 (7/21/2019), 213674 (7/21/2019), 213675 (7/21/2019), 213676 (7/21/2019), 213677 (7/21/2019), 213678 (7/21/2019), 213679 (7/21/2019), 213680 (7/21/2019), 213681 (7/21/2019), 213682 (7/21/2019), 213683 (7/21/2019), 213684 (7/21/2019), 213685 (7/21/2019), 213686 (7/24/2019), 213687 (7/24/2019), 213688 (7/24/2019), 213689 (7/24/2019), 213690 (7/24/2019), 213691 (7/24/2019), 213692 (7/24/2019), 213693 (7/24/2019), 213694 (7/24/2019), 213695 (7/24/2019), 213696 (7/24/2019), 213697 (7/24/2019), 213698 (7/24/2019), 213699 (7/24/2019), 213700 (7/24/2019), 213701 (7/24/2019), 213702 (7/24/2019), 213703 (7/24/2019), 213704 (7/24/2019), 213705 (7/24/2019), 213706 (7/24/2019), 213707 (7/24/2019), 213708 (7/24/2019), 213709 (7/24/2019), 213710 (7/24/2019), 213712 (7/24/2019), 213713 (7/24/2019), 213714 (7/24/2019), 213715 (7/24/2019), 213716 (7/24/2019), 213717 (7/24/2019), 213718 (7/24/2019), 213719 (7/24/2019), 213720 (7/24/2019), 213721 (7/24/2019), 213722 (7/24/2019), 213723 (7/24/2019), 213724 (7/24/2019), 213726 (7/24/2019), 213727 (7/24/2019), 213728 (7/24/2019), 213730 (7/24/2019), 213731 (7/24/2019), 213732 (7/24/2019), 213733 (7/24/2019), 213734 (7/24/2019), 213735 (7/24/2019), 213736 (7/24/2019), 213737 (7/24/2019), 213738 (7/24/2019), 213739 (7/24/2019), 213740 (7/24/2019), 213741 (7/24/2019), 213742 (7/24/2019), 213743 (7/24/2019), 213744 (7/24/2019), 213745 (7/24/2019), 213746 (7/25/2019), 213747 (7/25/2019), 213748 (7/25/2019), 213749 (7/25/2019), 213750 (7/25/2019), 213751 (7/25/2019), 213752 (7/25/2019), 213753 (7/25/2019), 213754 (7/25/2019), 213755 (7/25/2019), 213756 (7/25/2019), 213757 (7/25/2019), 213758 (7/25/2019), 213759 (7/25/2019), 213760 (7/25/2019), 213761 (7/25/2019), 213762 (7/25/2019), 213763 (7/25/2019), 213764 (7/25/2019), 213765 (7/25/2019), 213766 (7/25/2019), 213767 (7/25/2019), 213768 (7/25/2019), 213769 (7/25/2019), 213770 (7/25/2019), 213771 (7/25/2019), 213772 (7/25/2019), 213773 (7/25/2019), 213774 (7/25/2019), 213775 (7/25/2019), 213776 (7/25/2019), 213777 (7/25/2019), 213778 (7/25/2019), 213779 (7/25/2019), 213780 (7/25/2019), 213782 (7/25/2019), 213783 (7/25/2019), 213784 (7/25/2019), 213785 (7/25/2019), 213786 (7/25/2019), 213787 (7/25/2019), 213788 (7/25/2019), 213789 (7/25/2019), 213790 (7/25/2019), 213791 (7/25/2019), 213792 (7/25/2019), 213793 (7/25/2019), 213794 (7/25/2019), 213795 (7/25/2019), 213796 (7/25/2019), 213797 (7/25/2019), 213798 (7/25/2019), 213799 (7/25/2019), 213800 (7/25/2019), 213801 (7/25/2019), 213802 (7/25/2019), 213803 (7/25/2019), 213804 (7/25/2019), 213805 (7/25/2019), 213806 (7/25/2019), 213807 (7/25/2019), 213808 (7/25/2019), 213809 (7/25/2019), 213810 (7/25/2019), 213811 (7/25/2019), 213812 (7/25/2019), 213813 (7/25/2019), 213814 (7/25/2019), 213815 (7/25/2019), 213816 (7/26/2019), 213817 (7/26/2019), 213818 (7/26/2019), 213819 (7/26/2019), 213820 (7/26/2019), 213821 (7/26/2019), 213822 (7/26/2019), 213823 (7/26/2019), 213824 (7/26/2019), 213826 (7/26/2019), 213827 (7/26/2019), 213828 (7/26/2019), 213830 (7/26/2019), 213831 (7/26/2019), 213832 (7/26/2019), 213833 (7/26/2019), 213834 (7/26/2019), 213835 (7/26/2019), 213836 (7/26/2019), 213837 (7/26/2019), 213838 (7/26/2019), 213839 (7/26/2019), 213840 (7/26/2019), 213841 (7/26/2019), 213843 (7/26/2019), 213844 (7/26/2019), 213845 (7/26/2019), 213846 (7/26/2019), 213847 (7/26/2019), 213848 (7/26/2019), 213849 (7/26/2019), 213850 (7/26/2019), 213851 (7/26/2019), 213852 (7/26/2019), 213853 (7/26/2019), 213854 (7/26/2019), 213855 (7/26/2019), 213856 (7/26/2019), 213857 (7/26/2019), 213858 (7/26/2019), 213860 (7/26/2019), 213861 (7/26/2019), 213862 (7/26/2019), 213863 (7/26/2019), 213864 (7/26/2019), 213865 (7/26/2019), 213866 (7/26/2019), 213867 (7/26/2019), 213868 (7/26/2019), 213869 (7/26/2019), 213870 (7/26/2019), 213871 (7/26/2019), 213872 (7/26/2019), 213874 (7/26/2019), 213875 (7/26/2019), 213876 (7/26/2019), 213877 (7/26/2019), 213878 (7/26/2019), 213879 (7/26/2019), 213880 (7/26/2019), 213881 (7/26/2019), 213882 (7/26/2019), 213883 (7/26/2019), 213884 (7/26/2019), 213885 (7/26/2019), 213886 (7/27/2019), 213887 (7/27/2019), 213888 (7/27/2019), 213889 (7/27/2019), 213890 (7/27/2019), 213892 (7/27/2019), 213893 (7/27/2019), 213894 (7/27/2019), 213895 (7/27/2019), 213897 (7/27/2019), 213899 (7/27/2019), 213900 (7/27/2019), 213903 (7/27/2019), 213905 (7/27/2019), 213907 (7/27/2019), 213908 (7/27/2019), 213909 (7/27/2019), 213911 (7/27/2019), 213912 (7/27/2019), 213913 (7/27/2019), 213914 (7/27/2019), 213915 (7/27/2019), 213916 (7/27/2019), 213917 (7/27/2019), 213918 (7/27/2019), 213920 (7/27/2019), 213922 (7/27/2019), 213923 (7/27/2019), 213924 (7/27/2019), 213925 (7/27/2019), 213926 (7/27/2019), 213927 (7/27/2019), 213928 (7/27/2019), 213929 (7/27/2019), 213930 (7/27/2019), 213931 (7/27/2019), 213933 (7/27/2019), 213934 (7/27/2019), 213935 (7/27/2019), 213936 (7/27/2019), 213939 (7/27/2019), 213940 (7/27/2019), 213941 (7/27/2019), 213942 (7/27/2019), 213944 (7/27/2019), 213945 (7/27/2019), 213947 (7/27/2019), 213948 (7/27/2019), 213949 (7/27/2019), 213950 (7/27/2019), 213951 (7/27/2019), 213952 (7/27/2019), 213953 (7/27/2019), 213954 (7/27/2019), 213955 (7/27/2019), 213956 (7/28/2019), 213957 (7/28/2019), 213958 (7/28/2019), 213959 (7/28/2019), 213960 (7/28/2019), 213962 (7/28/2019), 213963 (7/28/2019), 213964 (7/28/2019), 213965 (7/28/2019), 213966 (7/28/2019), 213967 (7/28/2019), 213968 (7/28/2019), 213969 (7/28/2019), 213970 (7/28/2019), 213971 (7/28/2019), 213973 (7/28/2019), 213974 (7/28/2019), 213977 (7/28/2019), 213978 (7/28/2019), 213980 (7/28/2019), 213981 (7/28/2019), 213984 (7/28/2019), 213985 (7/28/2019), 213986 (7/28/2019), 213988 (7/28/2019), 213989 (7/28/2019), 213990 (7/28/2019), 213991 (7/28/2019), 213992 (7/28/2019), 213993 (7/28/2019), 213995 (7/28/2019), 213996 (7/28/2019), 213997 (7/28/2019), 213998 (7/28/2019), 213999 (7/28/2019), 214000 (7/28/2019), 214001 (7/28/2019), 214002 (7/28/2019), 214003 (7/28/2019), 214004 (7/28/2019), 214005 (7/28/2019), 214006 (7/31/2019), 214007 (7/31/2019), 214008 (7/31/2019), 214009 (7/31/2019), 214010 (7/31/2019), 214011 (7/31/2019), 214012 (7/31/2019), 214013 (7/31/2019), 214014 (7/31/2019), 214016 (7/31/2019), 214017 (7/31/2019), 214018 (7/31/2019), 214019 (7/31/2019), 214021 (7/31/2019), 214022 (7/31/2019), 214023 (7/31/2019), 214024 (7/31/2019), 214025 (7/31/2019), 214026 (7/31/2019), 214027 (7/31/2019), 214028 (7/31/2019), 214031 (7/31/2019), 214033 (7/31/2019), 214034 (7/31/2019), 214035 (7/31/2019), 214037 (7/31/2019), 214038 (7/31/2019), 214039 (7/31/2019), 214040 (7/31/2019), 214041 (7/31/2019), 214043 (7/31/2019), 214044 (7/31/2019), 214045 (7/31/2019), 214046 (7/31/2019), 214047 (7/31/2019), 214048 (7/31/2019), 214050 (7/31/2019), 214051 (7/31/2019), 214052 (7/31/2019), 214053 (7/31/2019), 214054 (7/31/2019), 214055 (7/31/2019), 214056 (7/31/2019), 214057 (7/31/2019), 214058 (7/31/2019), 214059 (7/31/2019), 214060 (7/31/2019), 214061 (7/31/2019), 214063 (7/31/2019), 214064 (7/31/2019), 214065 (7/31/2019), 214066 (7/31/2019), 214067 (7/31/2019), 214068 (7/31/2019), 214069 (7/31/2019), 214070 (7/31/2019), 214071 (7/31/2019), 214072 (7/31/2019), 214073 (7/31/2019), 214074 (7/31/2019), 214076 (7/31/2019), 214077 (7/31/2019), 214079 (7/31/2019), 214080 (7/31/2019), 214081 (7/31/2019), 214082 (7/31/2019), 214084 (7/31/2019), 214085 (7/31/2019)
Nationwide distribution to Alabama, Arizona, Connecticut, Delaware, Florida, Georgia, Hawaii, Illinois, Indiana, Iowa, Kansas, Louisiana, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Nebraska, New Hampshire, New Jersey, New Mexico, New York, North Carolina, Ohio, Oklahoma, Oregon, Pennsylvania, Puerto Rico, South Carolina, Tennessee, Texas, Vermont, Virginia, Washington, Wisconsin, Wyoming. International distribution to Argentina, Australia, Austria, Bahamas, Belgium, Bermuda, Brazil, Brunei Darussalam, Canada, Chile, China, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Finland, France, French Polynesia, Germany, Greece, Guam, Hong Kong, Hungary, India, Iran, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Latvia, Lebanon, , Lithuania, Luxembourg, Malaysia, Mexico, Monaco, Netherlands, New Caledonia, New Zealand, Norway, Oman, Poland, Portugal, Puerto Rico, Qatar, Reunion, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan R O C, Thailand, Tunisia, Turkey, U.S. - Virgin Islands, United Arab Emirates, United Kingdom.
404 devices