FDA Enforcement
Class II
Terminated
TactiCath Quartz Contact Force Ablation Catheter, Model Numbers: PN-004 065 (GTIN 07640157990033) and PN-004 075 (GTIN 07640157990040) - Product Usage: is a multi-electrode irrigated catheter with a deflectable tip designed to allow electrophysiological mapping of the heart and, when connected to an RF generator, to deliver radiofrequency current to the catheter tip electrode for the purpose of ablation.
Recall: Z-2541-2020
·
Reported July 15, 2020
Enforcement
- Recall Number
- Z-2541-2020
- Event ID
- 85775
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Abbott
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- July 15, 2020
- Initiation Date
- May 22, 2020
- Classification Date
- July 9, 2020
- Termination Date
- February 11, 2021
- Address
- 5050 Nathan Ln N, Plymouth, MN, 55442-3209, United States
Description
TactiCath Quartz Contact Force Ablation Catheter, Model Numbers: PN-004 065 (GTIN 07640157990033) and PN-004 075 (GTIN 07640157990040) - Product Usage: is a multi-electrode irrigated catheter with a deflectable tip designed to allow electrophysiological mapping of the heart and, when connected to an RF generator, to deliver radiofrequency current to the catheter tip electrode for the purpose of ablation.
Reason
High rate of reported complaints for the loss of contact force information while using the TactiCath Quartz Contact Force Ablation Catheter
Code Info
All batch numbers with an expiration date of April 25, 2020
Distribution
Worldwide distribution.
Quantity
84104 devices