FDA Enforcement Class II Terminated

PR4100 Microplate Reader US, Catalog Number 45591587 Product Usage: The PR4100 Microplate Reader is a 96 well absorbance reader for the measurement of light absorbance (optical density-OD) of liquid media. The instrument is intended to be used primarily for the in vitro diagnostic analysis of samples from the human body to obtain information oh physiological and pathological states.

Recall: Z-2541-2018 · Reported August 1, 2018

Enforcement

Recall Number
Z-2541-2018
Event ID
80403
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Bio-Rad Laboratories Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 1, 2018
Initiation Date
April 2, 2018
Classification Date
July 25, 2018
Address
6565 185th Ave NE, N/A, Redmond, WA, 98052-5039, United States

Description

PR4100 Microplate Reader US, Catalog Number 45591587 Product Usage: The PR4100 Microplate Reader is a 96 well absorbance reader for the measurement of light absorbance (optical density-OD) of liquid media. The instrument is intended to be used primarily for the in vitro diagnostic analysis of samples from the human body to obtain information oh physiological and pathological states.

Reason

The connector of the power supply was capable of fitting other components provided with the system. However, the voltage output of the reader power supply is incompatible with the other component and, if incorrectly connected, could result in irreparable damage to the system.

Code Info

UDI: 00847865014151 Serial Numbers: 140600072, 160900991, 161001953, 1112001938, 1112001941 , 1112004075 , 1112004079 , 1112004083 , 1112004084 , 1112004086 , 1112004088 , 1112004093 , 1201003822 , 1201003826 , 1201003827 , 1207001007 , 1207001015 , 1207001017 , 1207001018 , 1208001583 , 1208001584 , 1208001586 , 1208001590 , 1208005236 , 1208005244 , 1208005245 , 1211000978 , 1211000983 , 1211006938 , 1301003205 , 1301003212 , 1303000299 , 1303000300 , 1303000302 , 1303000305 , 1303007688 , 1306003195 , 1307006044 , 1307006045 , 1307006054 , 1307006055 , 1308000967 , 1308000968 , 1308000969 , 1308000971 , 1308000973 , 1308003338 , 1308003341 , 1308003345 , 1308003346 , 1308006670 , 1311008542 , 1311008544 , 1311008546 , 1311008555 , 1311008557 , 1311008562 , 1311008564 , 1311008565 , 1401002580 , 1401002583 , 1401002587 , 1401007785 , 1401007786 , 1401007789 , 1401007792 , 1401007793 , 1404000065 , 1404000066 , 1404000067 , 1404000068 , 1404002677 , 1404002684 , 1404002685 , 1405005640 , 1405005641 , 1405005645 , 1405005648 , 1406000064 , 1406000065 , 1406000066 , 1406000067 , 1406000068 , 1406000070 , 1406000071 , 1406000074 , 1406006883 , 1406006886 , 1407004144 , 1407004145 , 1407004147 , 1407004149 , 1407004151 , 1407004152 , 1410010410 , 1410010415 , 1410010416 , 1411010876 , 1411010877 , 1411010880 , 1412001714 , 1412001716 , 1412001721 , 1501001523 , 1501001526 , 1501001526 , 1501005823 , 1501005824 , 1501005825 , 1501005828 , 1501005829 , 1501005832 , 1506008687 , 1506008690 , 1506008695 , 1510005389 , 1510005393 , 1510005394 , 1510005395 , 1510005397 , 1510005398 , 1510005399 , 1510005414 , 1601001552 , 1601001553 , 1601001559 , 1601001561 , 1601001562 , 1601001563 , 1603001786 , 1603005407 , 1605000088 , 1609000992 , 1609000993 , 1609000994 , 1609000995 , 1609000996 , 1609001301 , 1609001302 , 1609001303 , 1609001304 , 1609001305 , 1609001306 , 1609009231 , 1610001955 , 1701001785 , 1701001788 , 1701004204 , 1701004210 , 1710010533 , 1710010539 , 13060003204 , 16303005411 , s1411010876

Distribution

US Nationwide Distribution inn the states of : AL, AR, AZ, CA, CO, CT, FL, GA, HI, IL, IN, KS, LA, MA, MD, MI, MN, MO, MS, NC, NJ, NY, OR, PA, SC, TN, TX, VA, VT, WA, WI, and Puerto Rico.

Quantity

153