Vue Optic Visualization Source, Flexor Vue Deflecting Endoscopic System, Endoscope, Fiber Optic, Supplied sterilized by ethylene oxide gas in peel-open packages
Enforcement
- Recall Number
- Z-2540-2017
- Event ID
- 77305
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Cook Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- June 21, 2017
- Initiation Date
- May 3, 2017
- Classification Date
- June 14, 2017
- Termination Date
- January 26, 2018
- Address
- 750 N Daniels Way, N/A, Bloomington, IN, 47404-9120, United States
Description
Vue Optic Visualization Source, Flexor Vue Deflecting Endoscopic System, Endoscope, Fiber Optic, Supplied sterilized by ethylene oxide gas in peel-open packages
Cook Inc. is initiating a voluntary recall of the Vue Optic Visualization Source and the Flexor Vue Deflecting Endoscopic System because it was determined that the current reprocessing instructions do not provide sufficient detailed information for the cleaning, disinfection, and sterilization of these products.
Vue Optic Visualization Source, Catalog Number: FVO-150, GPN: G25343 Flexor Vue Deflecting Endoscopic System Catalog Number: FV-090075-150, GPN: G50972 Catalog Number: FV-090045-150, GPN: G34306
Domestic: FL IN LA MS NC TX Foreign:Argentina, Australia, Brazil, Canada, Czech Republic, Egypt, Estonia, France, Germany, Israel, Italy, Netherlands, Norway, Paraguay, Poland, Portugal, Russian Federation, Spain, Switzerland, Tunisia, Turkey, United Kingdom, VA/DOD:None
1328 units