FDA Enforcement Class II Ongoing

Berchtold Chromophare F528 Surgical Lights. Product Usage: The medical lights are for usage in hospital treatment rooms. They provide localized lighting focused on the patient s body for detecting and treating disease, injury, and disability. The surgical lights may only be used in rooms used for medical purposes that have been equipped in accordance with VDE 0100-710 or IEC 60364-7-10 in accordance with the regulations.

Recall: Z-2539-2019 · Reported September 18, 2019

Enforcement

Recall Number
Z-2539-2019
Event ID
83355
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Stryker Communications
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
September 18, 2019
Initiation Date
March 22, 2019
Classification Date
September 12, 2019
Address
571 Silveron Blvd, Flower Mound, TX, 75028, United States

Description

Berchtold Chromophare F528 Surgical Lights. Product Usage: The medical lights are for usage in hospital treatment rooms. They provide localized lighting focused on the patient s body for detecting and treating disease, injury, and disability. The surgical lights may only be used in rooms used for medical purposes that have been equipped in accordance with VDE 0100-710 or IEC 60364-7-10 in accordance with the regulations.

Reason

The back cover of the light head was insufficiently assembled and could potentially fall off the light.

Code Info

Serial number ranges: 83180-69712678 to 83180-88970377;

Distribution

Distribution was nationwide, including to Puerto Rico. There was also government/military distribution. Foreign distribution was made to Argentina, Austria, Australia, Belgium, Bulgaria, Canada, China, Colombia, Denmark, France, Germany, Hong Kong, India, Indonesia, Italy, Japan, Kazakhstan, Korea, Kuwait, Lithuania, Mexico, Netherlands, Philippines, Poland, Russia, Saudi Arabia, South Africa, Spain, Sweden, Switzerland, Taiwan, United Arab Emirates, United Kingdom, Uruguay, and Vietnam.

Quantity

946 lights