FDA Enforcement Class II Terminated

AD-TECH Medical Instrument Corporation Drill Sleeve Guides DSG-6-3-090-2.4N utilizing raw material (RM0677) from the following lots from vendor 00-0001062., AD0021502, AD0021628. Product Usage: The drill sleeve guides are accessories to the Ad-Tech Cranial Drill Bits. The Cranial Drill Bits and accessories are intended to be used for drilling holes in the skull for neurological procedures, such as brain biopsy, brain contacting electrode and electrode accessory device placement. The DSG-6.3-090-2.4N drill sleeve guides are intended to be used with the 2.4mm Cranial Drill Bit diameter.

Recall: Z-2539-2018 · Reported August 1, 2018

Enforcement

Recall Number
Z-2539-2018
Event ID
80280
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Ad-Tech Medical Instrument Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 1, 2018
Initiation Date
May 2, 2018
Classification Date
July 25, 2018
Termination Date
February 8, 2019
Address
400 W Oakview Pkwy, N/A, Oak Creek, WI, 53154-7213, United States

Description

AD-TECH Medical Instrument Corporation Drill Sleeve Guides DSG-6-3-090-2.4N utilizing raw material (RM0677) from the following lots from vendor 00-0001062., AD0021502, AD0021628. Product Usage: The drill sleeve guides are accessories to the Ad-Tech Cranial Drill Bits. The Cranial Drill Bits and accessories are intended to be used for drilling holes in the skull for neurological procedures, such as brain biopsy, brain contacting electrode and electrode accessory device placement. The DSG-6.3-090-2.4N drill sleeve guides are intended to be used with the 2.4mm Cranial Drill Bit diameter.

Reason

The diameter of the drill sleeve guide raw material was found to be under tolerance, potentially resulting in the drill bit seizing in the guide during surgery.

Code Info

Catalog Number DSG-6.3-090-2.4N. Batch Number (Lot Number): 107104 (710220), 114390 (715001), 111423 (716180), 105176 (613121), 113775 (719190), 115847 (712211), 113776 (719190), 115000 (714201), 114724 (711101), 116389 (712121), 117146 (810110).

Distribution

US Nationwide distribution in the states of CA, FL, IL, MI, NH, NY, OH, OR, PA, SC, UT, WV. Worldwide distribution to Canada, Denmark, Germany, India, Israel, Japan, Russia, South Korea, Taiwan, United Kingdom.

Quantity

65 Total