Hookup Model/Catalog Number: 2-8-540CAS Product Description: Hookups provide the interface between an Advantage Plus" Automated Endoscope Reprocessor and Advantage Plus" Pass-Thru Automated Endoscope Reprocessor ( AER ) and endoscopes, which allow perfusion of liquids dispensed by the AER to contact and high-level disinfect the channels of flexible endoscopes.
Enforcement
- Recall Number
- Z-2538-2024
- Event ID
- 94987
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Steris Corporation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- August 14, 2024
- Initiation Date
- June 13, 2024
- Classification Date
- August 8, 2024
- Address
- 5960 Heisley Rd, Mentor, OH, 44060-1834, United States
Description
Hookup Model/Catalog Number: 2-8-540CAS Product Description: Hookups provide the interface between an Advantage Plus" Automated Endoscope Reprocessor and Advantage Plus" Pass-Thru Automated Endoscope Reprocessor ( AER ) and endoscopes, which allow perfusion of liquids dispensed by the AER to contact and high-level disinfect the channels of flexible endoscopes.
IFU for hookup were incorrectly reflected within hookup IFUs and online reference system, Hookup Lookup (HULU). Use of an incorrect hookup connection and/or parameter set may prevent an endoscope from being properly disinfected, which may present a risk of patient contamination or infection
Lot Code: UDI-DI 00677964086878;Lot Numbers 438165, 472172, 559354, 587523
Worldwide distribution - United States Nationwide and the countries of Canada, Belgium, Czech Republic, Denmark, Germany, Spain, Italy, Slovenia, Slovakia, Australia, Israel, Jordan, Kuwait, Namibia, Saudi Arabia, South Africa, Cote d Ivoire, Afghanistan, Albania, Belarus, Botswana, Kazakhstan, Malaysia, Qatar, Ukraine, United Arab Emirates, Nicaragua, and Nigeria.
6