FDA Enforcement Class II Terminated

VistaSolution, physiological signal transmitter. A software graphical user interface for use by healthcare professionals to display physiological data collected by the wireless remote monitoring system in healthcare settings. This is a secondary, adjunct patient monitor and is not intended to replace existing standard-of-care patient monitoring.

Recall: Z-2537-2020 · Reported July 15, 2020

Enforcement

Recall Number
Z-2537-2020
Event ID
85802
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Vitalconnect Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
July 15, 2020
Initiation Date
May 21, 2020
Classification Date
July 8, 2020
Termination Date
May 13, 2021
Address
224 Airport Pkwy Ste 300, San Jose, CA, 95110-1022, United States

Description

VistaSolution, physiological signal transmitter. A software graphical user interface for use by healthcare professionals to display physiological data collected by the wireless remote monitoring system in healthcare settings. This is a secondary, adjunct patient monitor and is not intended to replace existing standard-of-care patient monitoring.

Reason

A healthcare provider (HCP) noticed a patient had a lowered historic SpO2 reading that the software should have triggered a notification to the care team, but no notification was found. A software defect was encountered causing the lack of notification. The defect could result in exposing a patient to potential hypoxia for a short window of time. The firm is correcting the software to bring the performance back into compliance with specifications.

Code Info

Catalog number: 6200, Vista Solution version 2.1, UDI: 00850490007139

Distribution

U.S. Nationwide distribution including in the states of NJ, NY, VA, OUS: None

Quantity

7