FDA Enforcement Class II Terminated

Patient Positioning System with KRC2 controller using software versions 2.3.1 2.3.10 Product Usage: The patient positioning system is a SCARA designed robotic arm designed to position a patient for medical procedures prescribed by oncologists and others that require a high degree of accuracy and repeatability.

Recall: Z-2537-2018 · Reported August 1, 2018

Enforcement

Recall Number
Z-2537-2018
Event ID
80510
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Forte Automation Systems Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 1, 2018
Initiation Date
September 5, 2017
Classification Date
July 24, 2018
Termination Date
April 13, 2020
Address
8155 Burden Rd, N/A, Machesney Park, IL, 61115-8208, United States

Description

Patient Positioning System with KRC2 controller using software versions 2.3.1 2.3.10 Product Usage: The patient positioning system is a SCARA designed robotic arm designed to position a patient for medical procedures prescribed by oncologists and others that require a high degree of accuracy and repeatability.

Reason

Communications between the Patient Positioning System and the accuracy filter can periodically fail with no clear indication to the operator.

Code Info

PPS Serial Number Cabinet Type Software Version 7757-11 KRC2 2.3.3 7757-21 KRC2 2.3.3 7757-31 KRC2 2.3.3 7757-41 KRC2 2.3.3 7757-51 KRC2 2.3.8 7755-21 KRC2 2.3.3 7755-31 KRC2 2.3.3 7755-41 KRC2 2.3.3 7392-2 KRC2 2.3.3 7392-3 KRC2 2.3.3 7392-4 KRC2 2.3.3 7392-1 KRC2 2.3.8 7392-5 KRC2 2.3.8

Distribution

US in the state of California

Quantity

13