FDA Enforcement Class II Ongoing

Hookup Model/Catalog Number: 2-8-540 Product Description: Hookups provide the interface between an Advantage Plus" Automated Endoscope Reprocessor and Advantage Plus" Pass-Thru Automated Endoscope Reprocessor ( AER ) and endoscopes, which allow perfusion of liquids dispensed by the AER to contact and high-level disinfect the channels of flexible endoscopes.

Recall: Z-2536-2024 · Reported August 14, 2024

Enforcement

Recall Number
Z-2536-2024
Event ID
94987
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Steris Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 14, 2024
Initiation Date
June 13, 2024
Classification Date
August 8, 2024
Address
5960 Heisley Rd, Mentor, OH, 44060-1834, United States

Description

Hookup Model/Catalog Number: 2-8-540 Product Description: Hookups provide the interface between an Advantage Plus" Automated Endoscope Reprocessor and Advantage Plus" Pass-Thru Automated Endoscope Reprocessor ( AER ) and endoscopes, which allow perfusion of liquids dispensed by the AER to contact and high-level disinfect the channels of flexible endoscopes.

Reason

IFU for hookup were incorrectly reflected within hookup IFUs and online reference system, Hookup Lookup (HULU). Use of an incorrect hookup connection and/or parameter set may prevent an endoscope from being properly disinfected, which may present a risk of patient contamination or infection

Code Info

Lot Code: UDI-DI 00677964091223; Lot Numbers 404360, 415307, 421967, 423540, 425801, 427239, 435309, 436068, 447102, 516866, 523989, 539379, 540599, 542619, 544723, 550675, 554204, 560426, 587537, 588583, 594221, 604965, 61450

Distribution

Worldwide distribution - United States Nationwide and the countries of Canada, Belgium, Czech Republic, Denmark, Germany, Spain, Italy, Slovenia, Slovakia, Australia, Israel, Jordan, Kuwait, Namibia, Saudi Arabia, South Africa, Cote d Ivoire, Afghanistan, Albania, Belarus, Botswana, Kazakhstan, Malaysia, Qatar, Ukraine, United Arab Emirates, Nicaragua, and Nigeria.

Quantity

37