Arctic Sun Temperature Management System, Model 5000. The Arctic Sun¿ 5000 model is a non-invasive, thermal regulating system, indicated for monitoring and controlling patient temperature within a range of 32¿C to 38.5¿C (89.6¿F to 101.3¿F) in adult and pediatric patients of all ages.
Enforcement
- Recall Number
- Z-2536-2018
- Event ID
- 80375
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Bard Medical Division
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- August 1, 2018
- Initiation Date
- June 13, 2018
- Classification Date
- July 24, 2018
- Termination Date
- February 16, 2021
- Address
- 8195 Industrial Blvd NE, N/A, Covington, GA, 30014-1497, United States
Description
Arctic Sun Temperature Management System, Model 5000. The Arctic Sun¿ 5000 model is a non-invasive, thermal regulating system, indicated for monitoring and controlling patient temperature within a range of 32¿C to 38.5¿C (89.6¿F to 101.3¿F) in adult and pediatric patients of all ages.
An improper wiring connection on the chiller assembly may result in overheating of the connection and potential failure of certain components in the system.
Model #5000 Product codes: 5000-00-00, 5000-00-00L, 5000-00-00E, 5000-01-01, 5000-01-01L, 5000-01-02, 5000-01-03, 5000-01-04, 5000-01-05, 5000-00-06, 5000-01-07, 5000-01-08. All serial numbers are affected.
Domestic Distribution: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL,GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY. Worldwide Distribution: Canada, Austria, Bahrain, Belgium, Brazil, Brussels China, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Gibraltar, Greece, Guam, Hungary, Iran, Italy, Jordan, Korea, Luxemburg, Malaysia, Netherlands, Norway, Oman, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, UAE, UK, Vietnam.
5608 Units Total