FDA Enforcement
Class II
Terminated
DUFNER Instrumente GmbH-De Bakey Slim Needleholder- intended to hold a suturing needle for closing wounds during suturing and surgical procedures Reference Number: 18044-23
Recall: Z-2534-2020
·
Reported July 15, 2020
Enforcement
- Recall Number
- Z-2534-2020
- Event ID
- 85894
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Dufner Instrumente GmbH
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- July 15, 2020
- Initiation Date
- June 3, 2020
- Classification Date
- July 7, 2020
- Termination Date
- January 21, 2021
- Address
- Fohrenstr. 9, Tuttlingen, N/A, Germany
Description
DUFNER Instrumente GmbH-De Bakey Slim Needleholder- intended to hold a suturing needle for closing wounds during suturing and surgical procedures Reference Number: 18044-23
Reason
Tungsten carbide plate, which is intended to hold the sewing needle securely, can break when holding the sewing needle under pressure and result in in patient tissue injury
Code Info
Lot Numbers: 8170368 and 8180673
Distribution
IL
Quantity
30 units