FDA Enforcement
Class II
Terminated
Access Hi-Low Exam Tables, Model 6501-XX
Recall: Z-2531-2018
·
Reported August 1, 2018
Enforcement
- Recall Number
- Z-2531-2018
- Event ID
- 80164
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Brewer Company, LLC
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- August 1, 2018
- Initiation Date
- March 29, 2018
- Classification Date
- July 24, 2018
- Termination Date
- August 29, 2019
- Address
- N88 W13901 Main St, N/A, Menomonee Falls, WI, 53051-2330, United States
Description
Access Hi-Low Exam Tables, Model 6501-XX
Reason
One lot, the wire harness was over-crimped on some of the connectors that go to the two vertical side switches on the exam table.
Code Info
Model 6501-XX, Lot/Serial No. HL018786, HL018787, HL018788, HL018789, HL018790, HL018791, HL018797, HL018805
Distribution
Devices distributed in WI, TX, MN and IL.
Quantity
7 devices