FDA Enforcement
Class I
Ongoing
Hamilton-T1, REF: MSP161339/02, ESM Shielding Set, Spare Part for ventilator
Recall: Z-2530-2023
·
Reported September 27, 2023
Enforcement
- Recall Number
- Z-2530-2023
- Event ID
- 92873
- Classification
- Class I
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Hamilton Medical, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- September 27, 2023
- Initiation Date
- July 26, 2023
- Classification Date
- September 20, 2023
- Address
- 4655 Aircenter Cir, Reno, NV, 89502-5948, United States
Description
Hamilton-T1, REF: MSP161339/02, ESM Shielding Set, Spare Part for ventilator
Reason
Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.
Code Info
UDI: 07630002800761 / Affected Serial Numbers: 4050193, 4050210, 4050194 and 4016669
Distribution
Distribution US nationwide, including Puerto Rico. OUS: Not provided
Quantity
4 shielding set