FDA Enforcement
Class I
Ongoing
Hamilton C1/T1/MR1, REF: MSP161502/12, Control Board Spare part for ventilator
Recall: Z-2529-2023
·
Reported September 27, 2023
Enforcement
- Recall Number
- Z-2529-2023
- Event ID
- 92873
- Classification
- Class I
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Hamilton Medical, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- September 27, 2023
- Initiation Date
- July 26, 2023
- Classification Date
- September 20, 2023
- Address
- 4655 Aircenter Cir, Reno, NV, 89502-5948, United States
Description
Hamilton C1/T1/MR1, REF: MSP161502/12, Control Board Spare part for ventilator
Reason
Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.
Code Info
UDI: 07630002805247 / Affected Serial Numbers: 4001901, 4002139, 4003445, 4003567, 4003576, 4005595
Distribution
Distribution US nationwide, including Puerto Rico. OUS: Not provided
Quantity
6 units