FDA Enforcement Class I Ongoing

Hamilton-MR1, REF: 161010, Intensive Care Ventilator

Recall: Z-2528-2023 · Reported September 27, 2023

Enforcement

Recall Number
Z-2528-2023
Event ID
92873
Classification
Class I
Status
Ongoing
Product Type
Devices
Firm
Hamilton Medical, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
September 27, 2023
Initiation Date
July 26, 2023
Classification Date
September 20, 2023
Address
4655 Aircenter Cir, Reno, NV, 89502-5948, United States

Description

Hamilton-MR1, REF: 161010, Intensive Care Ventilator

Reason

Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.

Code Info

UDI: 07630002800761 & 07630002813556 / Affected Serial Numbers: 3422 3423 3424 3425 3426 3427 3428 3429 3469 3503 3510 3800

Distribution

Distribution US nationwide, including Puerto Rico. OUS: Not provided

Quantity

12 units