FDA Enforcement Class II Ongoing

LUMINOS Lotus Max -Intended to visualize anatomical structures by converting an x-ray pattern into a visible image Model Number: 11574100

Recall: Z-2527-2024 · Reported August 14, 2024

Enforcement

Recall Number
Z-2527-2024
Event ID
94948
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Siemens Medical Solutions USA, Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 14, 2024
Initiation Date
July 2, 2024
Classification Date
August 7, 2024
Address
40 Liberty Blvd, Malvern, PA, 19355-1418, United States

Description

LUMINOS Lotus Max -Intended to visualize anatomical structures by converting an x-ray pattern into a visible image Model Number: 11574100

Reason

The support arm may unintentionally lower resulting in injury to persons when they are under the display ceiling/wall suspension when positioning the displays

Code Info

UDI: (01)04056869269931(21) Serial Numbers: Serial Number 10251 10181 10278 10274 10115 10309 10024 10140 10073 10327 10091 10307 10253 10117 10311 10030 10194 10029 10063 10076 10173 10172 10170 10079 10234 10330

Distribution

US Nationwide distribution.

Quantity

26 units