FDA Enforcement Class II Terminated

MR Surgical Suite Option Product The MR Surgical Option when integrated with the GE Signa¿ 3.0T or 1.5T magnetic resonance scanner (previously cleared K041476) can be used to produce head and whole body magnetic resonance images that are high resolution high signal-to-noise ratio, with short scan times. The Signa¿ Excite MR system with Surgical Option is indicated for use as a diagnostic imaging device to produce axial sagittal, coronal and oblique images, spectroscopic images, and/or spectra, dynamic images of the internal structures and organs of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. The images produced by the Signa Excite system with Surgical Option reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis. The GE Signa Excite system with Surgical Option can provide MR images at various stages of a surgical procedure, and provides the capability to transfer a patient to the MR system table directly from a surgical table, or patient transport system. Images can be obtained using the MR scanner body coil, or local surface coils, and tile surface coils used with the integrated system may / can accommodate sterile draping for surgical procedures.

Recall: Z-2527-2018 · Reported August 1, 2018

Enforcement

Recall Number
Z-2527-2018
Event ID
80287
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
GE Medical Systems, LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 1, 2018
Initiation Date
March 30, 2018
Classification Date
July 23, 2018
Termination Date
March 29, 2024
Address
3200 N Grandview Blvd, N/A, Waukesha, WI, 53188-1693, United States

Description

MR Surgical Suite Option Product The MR Surgical Option when integrated with the GE Signa¿ 3.0T or 1.5T magnetic resonance scanner (previously cleared K041476) can be used to produce head and whole body magnetic resonance images that are high resolution high signal-to-noise ratio, with short scan times. The Signa¿ Excite MR system with Surgical Option is indicated for use as a diagnostic imaging device to produce axial sagittal, coronal and oblique images, spectroscopic images, and/or spectra, dynamic images of the internal structures and organs of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. The images produced by the Signa Excite system with Surgical Option reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis. The GE Signa Excite system with Surgical Option can provide MR images at various stages of a surgical procedure, and provides the capability to transfer a patient to the MR system table directly from a surgical table, or patient transport system. Images can be obtained using the MR scanner body coil, or local surface coils, and tile surface coils used with the integrated system may / can accommodate sterile draping for surgical procedures.

Reason

There is a potential safety issue with the table transfer release of the MR Surgical Suite System which may cause an unexpected loss of function of the patient transfer mechanism that allows the patient to be moved from the GE MR Surgical Table to the Maquet Magnus Surgical Table.

Code Info

Model Numbers M8074SS, M8053SS, M0074SS, M0001SS, 5265300-7, 5265300-8. Serial Number (System ID Number): 00000000000002 (602406IMR), 4295404 (4295404MR11), N/A (415UCBHMR3), N/A (720848MR4), 00000304124MR1 (904202MR8), N/A (309655IMRI), 4425007 (913588NMRW), 00000000UA0117 (616267IMRI), 00000000000003 (507284NMRV), 4311497 (MR229054), 4510086 (212636SIGNA3T), 00000000UA0116 (330344MR750W), UA0326 (614293MR9), N/A (614RMHTMR1), 4371613 (412692MR750W), N/A (214820ITABLE), 4250114 (409772MR6), 4550641 (4550641MR11), 4482037 (757388750WMR), 00000012228M63 (608263MR4IOR), 4356151 (414805IMR1), 00000299229MR5 (AH5824MR01), 00000012051M69 (AH5824SS01), 00000021158M61 (MRHM0885), 00000014131M67 (MRHM0323), 00000018480M64 (MR307144MR6), 4565719 (082427110075), 00000013254M68 (082445160014), 000000PTST0008 (082427040101), 000000PTST0007 (082427050024), 4596631 (082427310113), 4409769 (082427140133), 000000PTST0003 (CS1020MR01), DUMFMI60885001 (CS1006MR03), 00001309FMM07L (T4185506), 00000021695M62 (M4194477), 00000021695M62 (X41944020), 00000305145MR5 (UC2547MR01), 4320023 (083027822002316), 4508956 (A5381507), 00000010003M62 (EM0293), 00001219FMM0H1 (EM0105), 00000017217M61 (EM0105), 000000PTST0010 (YM3833), 00000000000004 (YM1733), 2937140 (0850270414), 4548529 (KW1008MR08), 000000PTST0006 (ZA2169MR01), 00001251FMM044 (34368MRS01), 4344097 (34418MRS01), 4605737 (N/A), 000000PTST0005 (10590MRS01), DUMFMI60885002 (10590MRS02), 4569734 (00132MRS01)

Distribution

US Nationwide distribution to AZ, CA, CO, FL, IL, KS, MI, MN, NY, OH, PA, TX, UT, VA, WI. Worldwide distribution to Austria, Brazil, China, Czech Republic, France, Germany, India, Italy, Japan, Korea, Kuwait, South Africa, Turkey, Turkey, United Arab Emirates, United Kingdom.

Quantity

54 units total