FDA Enforcement Class I Ongoing

Hamilton-T1, REF: 161006, Intensive Care and Transport Ventilator

Recall: Z-2526-2023 · Reported September 27, 2023

Enforcement

Recall Number
Z-2526-2023
Event ID
92873
Classification
Class I
Status
Ongoing
Product Type
Devices
Firm
Hamilton Medical, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
September 27, 2023
Initiation Date
July 26, 2023
Classification Date
September 20, 2023
Address
4655 Aircenter Cir, Reno, NV, 89502-5948, United States

Description

Hamilton-T1, REF: 161006, Intensive Care and Transport Ventilator

Reason

Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.

Code Info

UDI: 07630002801850 & 07630002813532 / Affected Serial Numbers: 15929 17600 17661 17677 17698 17699 17741 17850 17877 17913 17925 17955 17957 17969 17975 17976 17982 17985 17986 17991 17993 17995 18009 18166 18183 18189 18191 18199 18224 18244 18251 18255 18256 18260 18263 18272 18273 18276 18288 18289 18305 18308 18333 18340 18349 18360 18363 18377 18378 18383 18414 18415 18419 18441 18451 18458 18461 18502 18537 18544 18550 18778 18784 19237 19290 19291 19292 19293 19294 19298 19304 19305 19470 19483 19490 26117

Distribution

Distribution US nationwide, including Puerto Rico. OUS: Not provided

Quantity

76 units