FDA Enforcement Class II Terminated

uCT 760 System, Computed Tomography X-ray System

Recall: Z-2522-2020 · Reported July 15, 2020

Enforcement

Recall Number
Z-2522-2020
Event ID
85804
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Shanghai United Imaging Healthcare Co., Ltd.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
July 15, 2020
Initiation Date
May 8, 2020
Classification Date
July 6, 2020
Termination Date
January 20, 2022
Address
No. 2258 Chengbei Rd, Jiading Ind., Shanghai, N/A, China

Description

uCT 760 System, Computed Tomography X-ray System

Reason

Two issues were identified with the computed tomography x-ray system including an intermittent issue which may affect image data leading to artifacts; and intermittent scout scanning interruption due to occasional angle signal drift. This may result in the need for rescanning of the patient with the patient receiving an additional dose of radiation and possible use of additional contrast media.

Code Info

Model uCT 760; Serial Numbers: 600229, 600230, 600232, 600271 with software version R001.11.0.729741-Re-1100, and serial number 600282 with software version R001.11.0.785930-Re-1150.

Distribution

Distributed to two customers. Both customers located in Houston, Texas.

Quantity

5 units