FDA Enforcement
Class II
Terminated
uCT 760 System, Computed Tomography X-ray System
Recall: Z-2522-2020
·
Reported July 15, 2020
Enforcement
- Recall Number
- Z-2522-2020
- Event ID
- 85804
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Shanghai United Imaging Healthcare Co., Ltd.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- July 15, 2020
- Initiation Date
- May 8, 2020
- Classification Date
- July 6, 2020
- Termination Date
- January 20, 2022
- Address
- No. 2258 Chengbei Rd, Jiading Ind., Shanghai, N/A, China
Description
uCT 760 System, Computed Tomography X-ray System
Reason
Two issues were identified with the computed tomography x-ray system including an intermittent issue which may affect image data leading to artifacts; and intermittent scout scanning interruption due to occasional angle signal drift. This may result in the need for rescanning of the patient with the patient receiving an additional dose of radiation and possible use of additional contrast media.
Code Info
Model uCT 760; Serial Numbers: 600229, 600230, 600232, 600271 with software version R001.11.0.729741-Re-1100, and serial number 600282 with software version R001.11.0.785930-Re-1150.
Distribution
Distributed to two customers. Both customers located in Houston, Texas.
Quantity
5 units