FDA Enforcement Class II Ongoing

Brand Name: HeartMate Product Name: HeartMate 3" System Controller Model/Catalog Number: 106531US Software Version: N/A Product Description: The HeartMate 3" System Controller acts as the central power and communication hub for the HeartMate 3 LVAS. It passes power from the Power Module, the Mobile Power Unit, Lithium-Ion Batteries, or its own integrated emergency backup supply, down to the LVAD via the Driveline. The HeartMate 3 System Controller constantly monitors system performance through communication with the implanted LVAD and Controller internal measurements and alerts the user to any alarm conditions by activating membrane panel LEDs and integrated audio annunciators. Further information on alarm conditions as well as system status can be attained by the user from the front panel LCD on the System Controller. When connected to a HeartMate Touch" Communication System, the System Controller sends information regarding the System Controller and Pump Status once per second to provide additional information to the user. This link also allows the clinician to set new patient operating parameters (e.g. pump speed) and provides a link for downloading trend and/or event recorder data. Component: Yes, HeartMate 3 Left Ventricular System (LVAS) Kits

Recall: Z-2519-2026 · Reported June 24, 2026

Enforcement

Recall Number
Z-2519-2026
Event ID
99023
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Thoratec LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 24, 2026
Initiation Date
May 13, 2026
Classification Date
June 18, 2026
Address
6035 Stoneridge Dr, Pleasanton, CA, 94588-3270, United States

Description

Brand Name: HeartMate Product Name: HeartMate 3" System Controller Model/Catalog Number: 106531US Software Version: N/A Product Description: The HeartMate 3" System Controller acts as the central power and communication hub for the HeartMate 3 LVAS. It passes power from the Power Module, the Mobile Power Unit, Lithium-Ion Batteries, or its own integrated emergency backup supply, down to the LVAD via the Driveline. The HeartMate 3 System Controller constantly monitors system performance through communication with the implanted LVAD and Controller internal measurements and alerts the user to any alarm conditions by activating membrane panel LEDs and integrated audio annunciators. Further information on alarm conditions as well as system status can be attained by the user from the front panel LCD on the System Controller. When connected to a HeartMate Touch" Communication System, the System Controller sends information regarding the System Controller and Pump Status once per second to provide additional information to the user. This link also allows the clinician to set new patient operating parameters (e.g. pump speed) and provides a link for downloading trend and/or event recorder data. Component: Yes, HeartMate 3 Left Ventricular System (LVAS) Kits

Reason

Due to 11 volt Backup Battery failures.

Code Info

Lot Code: Model Number:106531US UDI/GTIN: 05415067039446 Serial Numbers Distributed in US: See Attachment E for serial numbers distributed.

Distribution

Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, DC, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI and the countries of Canada, Germany, Japan, Netherlands, Poland, Spain.

Quantity

694