BD Connecta" BD Luer-Lok" 360 REF 394910 UDI-DI code: 00382903949106 BD Connecta" Stopcocks are used to administer fluids or medication through one or two different ports via an IV cannula or extension tube. The ports can be used to sample blood or for hemodynamic monitoring. When used with IV lipid nutritional products, the stopcock device can be used for up to 24 hours.
Enforcement
- Recall Number
- Z-2514-2026
- Event ID
- 98855
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- BD SWITZERLAND SARL
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- June 24, 2026
- Initiation Date
- May 21, 2026
- Classification Date
- June 18, 2026
- Address
- Terre Bonne Park A4, Route De Crassier 17, Eysins, N/A, Switzerland
Description
BD Connecta" BD Luer-Lok" 360 REF 394910 UDI-DI code: 00382903949106 BD Connecta" Stopcocks are used to administer fluids or medication through one or two different ports via an IV cannula or extension tube. The ports can be used to sample blood or for hemodynamic monitoring. When used with IV lipid nutritional products, the stopcock device can be used for up to 24 hours.
Due to stopcock issues when using the plastic luer-lok syringe, the connection at the Port C may over thread or continue to spin rather than tightening securely. This behavior has led some clinical users to perceive the connection as unstable.
Catalog/Model Number: 394910 UDI-DI code: 00382903949106 Lot Numbers: 4277968 4277973 4282451 4282461 4289801 4290582 4292904 4295848 4299949 4303841 4305108 4306700 5129635 5129642 5129649 5129650 5129651 5139250 5139251 5139252 5150181 5153940 5164061 5164062 5178078 5181442 5183810 5183816 5183824 5183828 5218287 5218291 5218293 5218299 5218302 5241476 5247729 5247731 5251098 5275601 5275602
U.S.: AK, AL, AR, AZ, CA, CO, DC, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, and WV. O.U.S.: Japan
2,465,500 units