FDA Enforcement Class II Ongoing

NobelBiocare Tri-Lobe-compatible WP X 6.0mm Engaging Titanium Base, REF: 9000662

Recall: Z-2514-2025 · Reported September 10, 2025

Enforcement

Recall Number
Z-2514-2025
Event ID
97239
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Preat Corp
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
September 10, 2025
Initiation Date
June 25, 2025
Classification Date
September 3, 2025
Address
2625 Skyway Dr Ste B, N/A, Santa Maria, CA, 93455-1405, United States

Description

NobelBiocare Tri-Lobe-compatible WP X 6.0mm Engaging Titanium Base, REF: 9000662

Reason

Engaging titanium base was manufactured 180 degrees around the post from where it should be located in relation to the connection geometry; if the flat is on the wrong side of the post, the crown will not fit onto the abutment in the correct orientation, resulting in restoration delay.

Code Info

UDI-DI: 00842092169599. Lot: 277097

Distribution

US Nationwide distribution in the states of NY, FL, PA, CA, OH.

Quantity

11