FDA Enforcement Class I Ongoing

Proclaim Plus 7 Implantable Pulse Generator REF 3672 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more extensions or leads with up to 16 electrodes total. It is powered by a hermetically sealed battery within a titanium case and uses microelectronic circuitry to generate constant-current electrical stimulation. The IPG can deliver stimulation with a single program or with multiple programs. Each program can provide stimulation to a single anatomical area or to multiple areas. The IPG communicates wirelessly with system programmers and controllers, and the IPGs are available in small and large sizes to accommodate different power needs. The Proclaim" XR and Proclaim" Plus IPGs are used for Spinal Cord Stimulation (SCS). The Proclaim" DRG IPG is used for Dorsal Root Ganglion (DRG) stimulation. The Infinity" IPG is used for Deep Brain Stimulation (DBS).

Recall: Z-2508-2023 · Reported September 20, 2023

Enforcement

Recall Number
Z-2508-2023
Event ID
92688
Classification
Class I
Status
Ongoing
Product Type
Devices
Firm
Abbott Medical
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
September 20, 2023
Initiation Date
July 18, 2023
Classification Date
September 11, 2023
Address
6901 Preston Rd, Plano, TX, 75024-2508, United States

Description

Proclaim Plus 7 Implantable Pulse Generator REF 3672 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more extensions or leads with up to 16 electrodes total. It is powered by a hermetically sealed battery within a titanium case and uses microelectronic circuitry to generate constant-current electrical stimulation. The IPG can deliver stimulation with a single program or with multiple programs. Each program can provide stimulation to a single anatomical area or to multiple areas. The IPG communicates wirelessly with system programmers and controllers, and the IPGs are available in small and large sizes to accommodate different power needs. The Proclaim" XR and Proclaim" Plus IPGs are used for Spinal Cord Stimulation (SCS). The Proclaim" DRG IPG is used for Dorsal Root Ganglion (DRG) stimulation. The Infinity" IPG is used for Deep Brain Stimulation (DBS).

Reason

Firm has received complaints from patients who are unable to exit MRI (Magnetic Resonance Imaging) mode on their IPGs (Implantable Pulse Generators).

Code Info

Model Number: 3672 UDI-DI Codes: 05415067046406 Serial Numbers: US CBW328.1 CCA167.1 CCA177.1 CCU613.1 CCV760.1 CCX700.1 CCY890.1 CDR307.1 CDW220.1 CEC548.1 CEC719.1 CEE095.1 CEH111.1 CEH156.1 CEH200.1 CEH272.1 CEH293.1 CEH334.1 CEH964.1 CEH975.1 CEH993.1 CEK127.1 CEK258.1 CEK265.1 CEK271.1 CEK310.1 CEK432.1 CEM232.1 CEM340.1 CEM349.1 CEM375.1 CEM383.1 CEM396.1 CEN429.1 CEN444.1 CEP494.1 CEP497.1 CEP537.1 CEP540.1 CEP541.1 CEP553.1 CEP554.1 CEP562.1 CEP585.1 CEP645.1 CEP658.1 CEP663.1 CEP671.1 CEP685.1 CEP725.1 CEP730.1 CEP838.1 CEP847.1 CEP867.1 CER393.1 CER395.1 CER462.1 CER482.1 CER509.1 CER576.1 CET134.1 CET147.1 CET298.1 CET336.1 CET343.1 CET649.1 CEY608.1 CEY648.1 CFL299.1 CFL403.1 CFL540.1 CFL582.1 CFP963.1 CFP987.1 CFP991.1 CFR079.1 CFR100.1 CFR121.1 CFR392.1 CFR446.1 CFR478.1 CFR585.1 CFU079.1 CFU082.1 CFU091.1 CFU097.1 CFU098.1 CFU099.1 CFU121.1 CFU126.1 CFU132.1 CFU145.1 CFU333.1 CFU401.1 CFU426.1 CFU441.1 CFU460.1 CFU467.1 CFU472.1 CFU483.1 CFU526.1 CFU547.1 CFU569.1 CFU646.1 CFU687.1 CFV457.1 CFV588.1 CFV718.1 CFX211.1 CGA013.1 CGA029.1 CGA065.1 CGA097.1 CGA145.1 CGA331.1 CGA365.1 CGA406.1 CGA426.1 CGA451.1 CGA491.1 CGA500.1 CGA515.1 CGA581.1 CGC898.1 CGC908.1 CGC916.1 CGC921.1 CGC926.1 CGC955.1 CGC957.1 CGD063.1 CGD070.1 CGD076.1 CGD085.1 CGD101.1 CGD111.1 CGD119.1 CGD123.1 CGD192.1 CGD193.1 CGD220.1 CGD231.1 CGD240.1 CGD248.1 CGD251.1 CGD277.1 CGD278.1 CGD283.1 CGD308.1 CGD342.1 CGD343.1 CGD362.1 CGD694.1 CGD719.1 CGD739.1 CGD777.1 CGD790.1 CGD832.1 CGD892.1 CGG890.1 CGH131.1 CGH175.1 CGH183.1 CGH199.1 CGH245.1 CGH346.1 CGH413.1 CGL108.1 CGL207.1 CGL321.1 CGL863.1 CGL866.1 CGL889.1 CGP051.1 CGP078.1 CGP326.1 CGP615.1 CGS914.1 CGW231.1 CHB011.1 CHE598.1 CHE845.1 CHF161.1 CHF178.1 CHH717.1 CHH744.1 CHH773.1 CHJ032.1 CHK331.1 CHK354.1 CHK434.1 CHK648.1 CHM211.1 CHM389.1 CHM509.1 CHP603.1 CHP605.1 CHP621.1 CHR155.1 CHR240.1 CHT217.1 CHT238.1 CHT240.1 CHT241.1 CHT243.1 CHT245.1 CHT272.1 CHT356.1 CHT361.1 CHT373.1 CHT469.1 CHT478.1 CHT539.1 CHT553.1 CHT561.1 CHT589.1 CHT609.1 CHT624.1 CHT758.1 CHT815.1 CHT834.1 CHV674.1 CHV680.1 CHV710.1 CHV713.1 CHV794.1 CHV901.1 CHV989.1 CHW217.1 CHW262.1 CHW263.1 CHW306.1 CHW350.1 CHW362.1 CHW526.1 CHW560.1 CHW590.1 CHW591.1 CHW640.1 CHZ471.1 CHZ563.1 CHZ576.1 CJA246.1 CJG129.1 CJG177.1 CJM932.1 CJM940.1 CJN773.1 CJN780.1 CJN879.1 CJN987.1 CJP047.1 CJW418.1 CJW475.1 CJW534.1 CJW548.1 CJW586.1 CJX925.1 CJX962.1 CJX982.1 CJX986.1 CJY012.1 CJY075.1 CJY111.1 CJY226.1 CJY238.1 CJY304.1 CJY323.1 CJY360.1 CJY409.1 CJZ061.1 CJZ103.1 CJZ135.1 CJZ157.1 CJZ224.1 CJZ241.1 CJZ253.1 CJZ256.1 CJZ283.1 CJZ293.1 CJZ300.1 CJZ330.1 CJZ420.1 CJZ444.1 CJZ450.1 CKB402.1 CKB428.1 CKB439.1 CKB447.1 CKB487.1 CKB507.1 CKB608.1 CKB624.1 CKB654.1 CKB696.1 CKB725.1 CKB767.1 CKB887.1 CKD072.1 CKF569.1 CKF587.1 CKG394.1 CKG550.1 CKG586.1 CKG607.1 CKG915.1 CKG954.1 CKG956.1 CKG962.1 CKH467.1 CKH606.1 CKH692.1 CKH719.1 CKH736.1 CKH765.1 CKH774.1 CKH786.1 CKH815.1 CKH859.1 CKH923.1 CKL195.1 CKL217.1 CKL236.1 CKL295.1 CKL345.1 CKL464.1 CKL487.1 CKP251.1 CKP255.1 CKP277.1 CKP291.1 CKP393.1 CKP435.1 CKP483.1 CKP495.1 CKP563.1 CKP656.1 CKP678.1 CKP682.1 CKP717.1 CKP733.1 CKP803.1 CKP939.1 CKS483.1 CKS519.1 CKS531.1 CKS552.1 CKS588.1 CKS590.1 CKS616.1 CKS641.1 CKS662.1 CKS673.1 CKS684.1 CKS692.1 CKS698.1 CKS706.1 CKS827.1 CKS856.1 CKS869.1 CKS954.1 CKS986.1 CKS992.1 CKS994.1 CKS997.1 CKT159.1 CKY723.1 CKY781.1 CLC604.1 CLD150.1 CLD293.1 CLF739.1 CLF855.1 CLF880.1 CLF888.1 CLF901.1 CLG005.1 CLG094.1 CLG124.1 CLG391.1 CLG393.1 CLG395.1 CLG396.1 CLG423.1 CLG457.1 CLG534.1 CLK164.1

Distribution

Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The countries of Algeria, Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, Colombia, Costa Rica, Czech Republic, Denmark, Ecuador, Estonia, Finland, France, Germany, Greece, Hungary, India, Iraq, Ireland, Israel, Italy, Japan, Korean, Kuwait, Lithuania, Luxembourg, Mexico, Netherlands, New Zealand, Norway, Pakistan, Poland, Portugal, Reunion, Russian Federation, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom.

Quantity

387 devices (US only)