FDA Enforcement
Class II
Terminated
CoLink Bone Graft Harvester, 8mm, Sterile, Disposable, REF G05 S1008 - Product Usage: is a general instrument intended for assisting surgeons in the manipulation of bone during surgical procedures.
Recall: Z-2506-2021
·
Reported September 29, 2021
Enforcement
- Recall Number
- Z-2506-2021
- Event ID
- 88510
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- In2bones USA, LLC
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- September 29, 2021
- Initiation Date
- August 18, 2021
- Classification Date
- September 22, 2021
- Termination Date
- August 8, 2023
- Address
- 6000 Poplar Ave Ste 115, Memphis, TN, 38119-3981, United States
Description
CoLink Bone Graft Harvester, 8mm, Sterile, Disposable, REF G05 S1008 - Product Usage: is a general instrument intended for assisting surgeons in the manipulation of bone during surgical procedures.
Reason
Complaints were received of CoLink Bone Graft Harvesters breaking at the outer tube weld site.
Code Info
Lot Numbers: 992921, 1018971, 1046581, and 1070081.
Distribution
US Nationwide distribution in the states of AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, LA, ME, MI, MN, MO, NC, NE, NJ, NM, NV, OH, OK, PA, SC, TN, TX, VA, and WI. There was no military/government distribution.
Quantity
835 units