FDA Enforcement Class II Ongoing

Philips Allura System configured with a 1 Phase UPS. Allura family (R1.x, R2.x). Labeled as the following with corresponding model numbers: 1. Allura Xper FD10, Model Number 722003 2. Allura Xper FD10/10, Model Number 722005 3. Allura Xper FD20, Model Number 722006 4. Allura Xper FD20 Biplane, Model Number 722008 5. Allura Xper FD10, Model Number 722010 6. Allura Xper FD10/10, Model Number 722011 7. Allura Xper FD20, Model Number 722012 8. Allura Xper FD20 Biplane, Model Number 722013 9. Allura Xper FD20 OR Table, Model Number 722023 10. Allura Xper FD10, Model Number 722026 11. Allura Xper FD10/10, Model Number 722027 12. Allura Xper FD20, Model Number 722028 13. Allura Xper FD20/10, Model Number 722029 14. Allura Xper FD20 OR Table, Model Number 722035 15. Allura Xper FD20/20, Model Number 722038 16. Allura Xper FD20/15, Model Number 722058 17. Allura Xper FD20/15 OR Table, Model Number 722059.

Recall: Z-2503-2025 · Reported September 10, 2025

Enforcement

Recall Number
Z-2503-2025
Event ID
97249
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
September 10, 2025
Initiation Date
August 1, 2025
Classification Date
September 3, 2025
Address
Veenpluis 4-6, N/A, Best, N/A, N/A, Netherlands

Description

Philips Allura System configured with a 1 Phase UPS. Allura family (R1.x, R2.x). Labeled as the following with corresponding model numbers: 1. Allura Xper FD10, Model Number 722003 2. Allura Xper FD10/10, Model Number 722005 3. Allura Xper FD20, Model Number 722006 4. Allura Xper FD20 Biplane, Model Number 722008 5. Allura Xper FD10, Model Number 722010 6. Allura Xper FD10/10, Model Number 722011 7. Allura Xper FD20, Model Number 722012 8. Allura Xper FD20 Biplane, Model Number 722013 9. Allura Xper FD20 OR Table, Model Number 722023 10. Allura Xper FD10, Model Number 722026 11. Allura Xper FD10/10, Model Number 722027 12. Allura Xper FD20, Model Number 722028 13. Allura Xper FD20/10, Model Number 722029 14. Allura Xper FD20 OR Table, Model Number 722035 15. Allura Xper FD20/20, Model Number 722038 16. Allura Xper FD20/15, Model Number 722058 17. Allura Xper FD20/15 OR Table, Model Number 722059.

Reason

Philips Azurion and Allura systems configured with a 1 Phase Uninterruptible Power Supply (UPS) have the potential of shutting down and/or not starting up, if power to the system is lost.

Code Info

1. Model Number/System Code (UDI); Serial Numbers: 722003 (No UDI); 1225, 789, 1138, 445, 740, 152, 1411. 2. Model Number/System Code (UDI); Serial Numbers: 722005 (No UDI); 35. 3. Model Number/System Code (UDI); Serial Numbers: 722006 (No UDI); 168, 2198, 2184, 2606, 195, 161, 1956, 2256, 2070, 1502, 376, 1096, 1731. 4. Model Number/System Code (UDI); Serial Numbers: 722008 (No UDI); 246, 2470. 5. Model Number/System Code (UDI); Serial Numbers: 722010 (00884838059030); 911, 1180, 936, 338, 660, 411, 992, 1210, 873, 1028, 921. 6. Model Number/System Code (UDI); Serial Numbers: 722011 (00884838059047); 308. 7. Model Number/System Code (UDI); Serial Numbers: 722012 (00884838059054); 525, 823, 1141, 1078, 1811, 2092, 1641, 114, 2134, 1589, 1446, 1675, 1625, 1989, 2156, 455, 233, 2929, 2930, 749, 1080, 149, 854. 8. Model Number/System Code (UDI); Serial Numbers: 722013 (00884838059061); 580, 254, 78, 357, 428. 9. Model Number/System Code (UDI); Serial Numbers: 722023 (00884838059085); 164, 66. 10. Model Number/System Code (UDI); Serial Numbers: 722026 (00884838054189); 287, 825, 462, 192, 74, 191, 204, 155, 967, 1132, 205, 1246, 253, 156, 234, 233, 845, 122, 259, 185, 618, 531, 855, 235, 335, 1068, 843, 805, 646, 965, 258, 220, 602, 908, 221, 664, 109, 110, 116, 117, 115, 284, 389, 223, 694, 222, 436, 1192, 102, 914, 396, 704, 370, 201, 774, 696, 838, 344, 93, 1072, 601, 245, 1340, 604, 627, 190, 605, 935, 297, 98, 99, 100, 101, 544, 1129, 1268, 841, 1239, 1293, 648, 1038, 1231, 215, 683, 142, 1050, 224, 73, 608, 1051, 1052, 323, 324, 931, 265, 431, 654, 786, 617, 87, 990, 1101, 738, 1287, 75, 688, 151, 121, 78, 114, 79, 770, 909, 1000, 612, 319, 458, 381, 826, 716, 432, 244, 534, 807, 962, 986, 961, 638, 830, 759, 1115, 707, 271, 296, 257, 86, 926, 448, 536, 525, 682, 1218, 1273, 1156, 146, 343, 406, 722, 247, 970, 672, 824, 903, 1234, 1065, 673, 603, 888, 790, 758, 314, 304, 606, 263, 262, 457, 1222, 251, 441, 686, 1024, 1086, 992, 178, 1178, 435, 710, 592, 1245, 660, 345, 118, 1219, 333, 338, 360, 730, 551, 349, 727, 519, 133, 195, 196, 137, 528, 460, 529, 530, 126, 455, 542, 566, 978, 1201, 1202, 22, 277, 1317, 545, 723, 359, 846, 506, 197, 628, 1001, 119, 120, 857, 789, 481, 1138, 818, 734, 1137, 154, 313, 502. 11. Model Number/System Code (UDI); Serial Numbers: 722027 (00884838054196); 203, 123, 133, 284, 134, 222, 255, 106, 107, 234, 42, 179, 306, 307, 265, 267, 40, 332, 112, 305, 104, 231, 263, 342, 71, 351, 298, 116, 50, 29, 39, 49, 56, 152, 153, 202, 164, 160, 64, 99, 101, 254, 84, 297, 268, 77, 115, 114, 142, 30, 79, 38, 189, 300, 299, 247, 282, 70. 12. Model Number/System Code (UDI); Serial Numbers: 722028 (00884838054202); 596, 819, 1598, 316, 1946, 541, 565, 647, 674, 864, 710, 946, 949, 880, 1041, 447, 696, 166, 618, 313, 1214, 1846, 1978, 1649, 1256, 904, 242, 284, 1397, 1957, 207, 208, 526, 1677, 1756, 431, 2760, 601, 377, 1454, 1629, 275, 2933, 1340, 539, 356, 236, 237, 238, 422, 525, 350, 560, 1407, 272, 494, 80, 324, 282, 262, 1275, 406, 540, 1017, 840, 842, 136, 137, 793, 1130, 260, 636, 1446, 741, 768, 901, 1657, 1828, 1988, 638, 267, 654, 759, 1656, 1736, 1076, 929, 251, 989, 125, 211, 368, 499, 261, 1424, 1261, 1854, 924, 934, 1142, 1651, 774, 2174, 1463, 2846, 779, 1111, 519, 590, 453, 452, 680, 1074, 318, 11, 1191, 1212, 1211, 1269, 340, 481, 468, 317, 1774, 2502, 982, 339, 128, 655, 143, 1482, 1405, 140, 1448, 1413, 464, 2103, 139, 303, 476, 419, 521, 522, 1132, 580, 861, 1014, 256, 1266, 917, 1472, 2007, 2008, 1652, 1396, 1646, 1394, 254, 326, 1824, 343, 424, 435, 437, 543, 130, 987, 1187, 888, 273, 336, 375, 887, 533, 516, 449, 408, 1549, 376, 744, 1034, 420, 553, 110, 2261, 983, 627, 903, 1099, 1481, 363, 1690, 1233, 2206, 2310, 2506, 1331, 365, 203, 204, 1538, 625, 528, 310, 743, 445, 189, 292, 1466, 1200, 1374, 173, 161, 620, 1085, 361, 281, 475, 456, 1310, 315, 908, 672, 1522, 127, 132, 141, 145, 180, 1001, 463, 1484, 1835, 1244, 1022, 1004, 2173, 210, 1350, 1723, 1834, 2061, 2710, 591, 564, 407, 1751, 2284, 697, 364, 520, 2070, 582, 932, 930, 129, 218, 1483, 485, 176, 257, 1920, 962, 1809, 279, 459, 2196, 889, 556, 1149, 1095, 1870, 2253, 347, 909, 1498, 1650, 952, 160, 1569, 966, 306, 234, 213, 1202, 155, 222, 1122, 131, 133, 1855, 224, 490, 1570, 2249, 1869, 1447, 631, 637, 529, 493, 534, 491, 206, 1890, 205, 571, 197, 194, 195, 216, 217, 229, 227, 228, 305, 488, 1822, 1880, 209, 825, 1461, 2709, 2707, 579, 1634, 48, 1783, 498, 462, 1043, 142, 159, 280, 535, 2252, 158, 184, 185, 473, 489, 1826, 466, 671, 978, 402, 1499, 386, 1710, 1708, 1326, 1032, 454, 918, 1705, 914, 698, 1387, 666, 1995, 1889, 1983, 1568, 1980, 1845, 416, 394, 428, 1368, 426, 559, 684, 1333, 547, 550, 190, 973, 1068, 1069, 1630, 346, 1156, 632, 1567, 2407, 2422, 1428, 121, 333, 151, 372, 371, 911, 1633, 717, 1155, 265, 1343, 1439, 1600, 758, 942, 878, 1572, 1039, 763, 231, 663, 822, 1008, 430, 200, 1292, 1046, 728, 1088, 951, 440, 536, 1390, 1645, 827, 1378, 2372, 2373, 503, 2877, 2878, 2720, 853, 415, 1932, 9, 578, 2168, 635, 412, 575, 799, 766, 776, 434, 389, 380, 624, 119, 441, 1937, 105, 111, 687, 2076, 577, 954, 192, 1907, 810, 1128, 899, 1309, 472, 1112, 1307, 1308, 51, 767, 800, 826, 1021, 1235, 1329, 1462, 2314, 1135, 1134, 395, 1210, 2153, 29, 802, 1258, 398, 665, 328, 439, 1632, 322, 2112, 621, 226, 311, 1118, 611, 1120, 1604, 1602, 1601, 1603, 830, 1195, 352, 1268, 1125, 1879, 469, 775, 502, 2274, 113. 13. Model Number/System Code (UDI); Serial Numbers: 722029 (00884838054219); 91, 71, 110, 126, 21, 62, 88, 36, 100, 30, 63, 60, 94, 95, 56, 31, 107, 43. 14. Model Number/System Code (UDI); Serial Numbers: 722035(00884838054240); 85, 24, 101, 117, 21, 38, 177, 121, 183, 88, 57, 125, 119, 100, 165, 47, 52, 131, 124, 169, 46, 74, 30, 60, 59, 196, 216, 187, 34, 222, 72, 111, 128, 1, 96, 53, 160, 104, 191, 70, 154, 142, 134, 129, 65. 15. Model Number/System Code (UDI); Serial Numbers: 722038 (00884838054226); 39, 63, 42, 55, 78, 99, 60, 71, 115, 108, 67, 106, 118, 130, 49, 119, 126, 46, 37, 38, 48, 62, 35, 120, 41, 157, 96, 70, 137, 166, 34, 116, 72, 64. 16. Model Number/System Code (UDI); Serial Numbers: 722058 (00884838059115); 337, 22, 202, 71, 239, 186, 56, 24, 168, 129, 123, 259, 134, 38, 171, 264, 61, 99, 162, 65, 278, 277, 196, 23, 205, 212, 111, 49, 89, 150, 177, 60, 131, 377, 377, 206, 106, 223, 229, 288, 169, 125, 126, 320, 139, 213. 17. Model Number/System Code (UDI); Serial Numbers: 722059 (00884838059122); 7, 8.

Distribution

Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Colombia, Cyprus, Czech Republic, Dominican Republic, Ecuador, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Iran, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Korea, Republic of, Kosovo, Latvia, Lithuania, Macedonia, Maldives, Malta, Mexico, Moldova, Netherlands, New Caledonia, New Zealand, Nicaragua, Norway, Panama, Paraguay, Peru, Philippines, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Tunisia, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Viet Nam.

Quantity

1,006 units