Centricity Universal Viewer and Viewer 6.0 - Product Usage: Is a device that displays medical images (including mammograms) and data from various imaging sources. Includes the following Model Numbers: 1. Model Number: 2090255-001; 2. Model Number: 2088026-723; 3. Model Number: 2102676-001; 4. Model Number: 2104867-044.
Enforcement
- Recall Number
- Z-2503-2021
- Event ID
- 88526
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- GE Healthcare, LLC
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- September 29, 2021
- Initiation Date
- August 6, 2021
- Classification Date
- September 22, 2021
- Address
- 3000 N Grandview Blvd, Waukesha, WI, 53188-1615, United States
Description
Centricity Universal Viewer and Viewer 6.0 - Product Usage: Is a device that displays medical images (including mammograms) and data from various imaging sources. Includes the following Model Numbers: 1. Model Number: 2090255-001; 2. Model Number: 2088026-723; 3. Model Number: 2102676-001; 4. Model Number: 2104867-044.
Image acquisition failures and synchronization failure with the Centricity Enterprise Archive
1. Model Number: 2090255-001 (Centricity Universal Viewer with PACS-IW foundation software versions 6.0 through 6.0 SP7.1) System ID (UDI-DI if applicable): PMIO14194768, PMIO20084199, PACS-IW-231, CITRUSVAL20884363, CITRUSVAL20884364, CITRUSVAL20884365, PACS-IW-155, PRIMEHEAL02315681, PRIMEHEAL20501355, PARKVIEWC21200827 (840682103800), PARKVIEWC21200830 (840682103800), PACS-IWUVN2-441, QUEEN19630914, QUEEN19630915, QUEEN19630916T, QUEEN20553950, 203268UVSW, DELAWAREC01589220, PACS-IW-462UV, PACS-IW-462UVN1, PACS-IW-126, PACS-IW-1062, OFMC01819402, OFMC20791988, HGS01766912, HGS20584004 (840682103800), GA001PACTRL01, TRGI02180958, DRY01810490, DRY19670358, PACS-IW-101, RML20502096, PACS-IWUV-239, PACS-IW-00813, PHS01374810, 01722652SEAS, 02031921SEASN1, MARYLANDF17985489, MYRTLEHIL17713895, MYRTLEHIL01444483, MYRTLEHIL17713895, VT01766830, VT01977629, PACS-IW-2637, SRMC01982629, PACS-IWUV-568, 1-142022244, PACS-IWUV-1131, OGAA02373799NOD1, ALLG00558571 (840682103800), EXC02189547, EXC01584099, PACS-IW-574, PRIME HEAL19508102, PRIME HEAL19050922, PRIME HEAL19050921, PACS-IW-219, ANCL00558570, ANCL02157634, NSW01452088, NSW20080545, MARYLANDF17985489, PACS-IW-518UV (840682103800), MTGD02393053, MTGD00558454, PACS-IW-44, QUENTIN01970717, LAKESIDEM02256808, BMC12P00558644P1, BMC12T02267839, BMC12N01271417N1, PACS-IW-58, SMCO02039946PC (840682103800), BWH01301357, PACS-IW-349, PACS-IW-255, 1-357059471, CAROCOMMU02205222, CAROCOMMU01459013, CSPA01291589 (840682103800), CSPA02097167 (840682103800), PACS-IW-79, PACS-IW-93, 1-276423210, MFCI01448329, LSFH02032334, CORONAREG01300927, 951737IWUWWN1, CORONAREG02288447, 1-382753509, CITRUSVAL19618406, MGMC02288391, MGMC02398531, 1-473998727, BARRETTHO00558704, BARRETTHO17737529, FORTDUNCA01298556 (840682103800), FORTDUNCA02395799 (840682103800), 1-283597082, TTU01973509, AVHL02396030, PACS-IW-511, PACSIWTEST, PACS-IW-386UV (840682103800), PACS-IW-486-UV (840682103800), PACS-IW-486-UVN1 (840682103800), PACS-IW-390 (840682103800), PACS-IW-202TEST, PACS-IW-202, HZC14961014, AVHL02396026, AVHL02396027, AVHL02396028, AVHL02396029, PACS-IW-419, PMRM14768587UVProxy, HANNIBALR02179353 (840682103800), TESTHANNIBALR02255613 (840682103800), HANN16174584UVProxy (840682103800), PACS-IW-189 (840682103800), HARTSDALE14768195 (840682103800), HOA01451973 (840682103800), RML01448365 (840682103800), ICA01367198 (840682103800), ICA14776604 (840682103800), JKS01452133 (840682103800), JKS14776447UVSFNode (840682103800), JKS14776448UVProxy (840682103800), JKS14776449UVTest (840682103800), PACS-IW-181, GTEC01477448, GTEC01477447, GTEC15993728UVProxy3, 1-592001096 (840682103800), UHSD14767800 (840682103800), PACS-IW-551UVN1 (840682103800), PACS-IW-551UVN2 (840682103800), UV1-154550926, DHOL19603374, DHOL19603375, PACS-IW-551UV (840682103800), PACS-IW-424CTLR, PACS-IW-424UVN1, PACS-IW-424UVN2, SFCS19667347, SFCS15961473, SFCS15961473, MHOG01806096, PACS-IW-166, PACS-IW-166T, MCT02277843, MCT02278145, PACS-IW-565, PACS-IWT-565, WGH00558334CTRL, WGH02372377PROX, WGH02372375BACK, 1-464783615, PACS-IW-389, BVH00558643, PACS-IW-348UV (840682103800), NRVM19609993UVT (840682103800), NRVM19611189UVPROXY (840682103800), PACS-IW-441, PACS-IW-441TUVSW, PACS-IW-402, PACS-IW-318, MINNETOHE, PACS-IW-253, PACS-IW-581UV, PACS-IW-370, PACS-IW-UVT370, VISOU02345023, VISOU02288781, 1-218291079, 1-103108035, MFHO14767332, PALMDALEM00558700, PALMDALEM01810788, PACS-IW-454, PACS-IW-406, 20101378UBSW, ILCC01373579 (840682103800), ILCC14580319 (840682103800), 516473UVSW (840682103800), JOHNTMA19732280T (840682103800), 1-182669905, NCDC20279342, CENTERFOR01627896, CENTERFOR16023960UVProxy, PACS-IW-176, PACS-IW-176UVTEST, PACS-IW-405, PACS-IW-336, NASONHOSP14898835, RRA01814001, MYRTLEHIL01444483, PSPC01815255, 1-283596904, HOOO20091140, GZMG02182469CNTRLR (840682103800), GZMG02287985 (840682103800), GZMG02288300 (840682103800), PACS-IW-556UV, URSA20025013, PACS-IW-431, 1-598891598, PACS-IW-230, INTERCOAST17594527UVProxy, MDHC16204735UVProxy, 201392UVSW, 505368UVBSW (840682103800), NNAR14767799 (840682103800), PACS-IW-479, PACS-IW-557 (840682103800), URS20583582 (840682103800), PACS-IWUV-481, PACS-IW-476, PACSIW476UVTEST, GGMC02349137, 1-486091448, 1-486091475, 1-750710672, SUC00558667, 1-802428413, 954688UVBSW100, PACS-IW-1511, KDH217192544, ACP00558109, PACS-IW-31Test1, PACS-IW-31Test, ACP01613226, 1-219095008, OKOS14767330, 918277CVUWBSW, CCA20759776SF, PACS-IW-487, QTHBL14767810UVPROXY, QTHBL14767809UVSFBK, QTHBL14767808UVSF1KO, QTHBL14767807UVSFSIM, QTHBL14767806UVSFVAL, QTHBL02256595, VCM02255583T, VCM01819618, VCM02255586TSP, VMC02255587NSP, BOJC00558520, PACS-IW-56, PACS-IW-225, 1-375705893, DDNI14776603, PACS-IW-129, DISSFHW01, 1-644600677, PACS-IW-154UV, FLMC14777071, NEWYORK01291738, OANT02035061, PACS-IW-424CTLR, PACS-IW-424UVN1, PACS-IW-424UVN2, VBHCP20497227, PACS-IW-2635, PREMI02182462, PREMI02182462, NORTHWEST01804127, HCIT4271649UW, AH4325SV01, AH5291PC02, AP6264PC01, AP6469PC01, AH0116SV03, AH0106SV03, HCD4440494UV, HCIT4258559UV, HCIT7109403UV, HCIT4244609UV, HCIT4708563UV, HCIT4470563UV-CTRL, HCIT4261024UVCTRL, HCIT7113573UV, 519264PACSAW1 (840682103800), 416813UV (840682103800), 416813TESTRPACSUV (840682103800), HCIT4266470UW, HCIT4238540UV, HCIT4486723UV, E4167658, X314119802, X141385617, A5485262 / A5485266 / A5485267 / A5485269 / A5485270 / A5485271, A9190177, M41980107, M4487535, M44997011, HC0399UV01, CC4105PC02, PC1695SV01, 786787SV02, HBON01SV02, HC4105PC01, 509940UV01, HC0529PC01, HC0529PC01, 233638UV01, HC0245PC02, 600791PC02, 427990UV01, PC1961SV02, 531344PC01, 434487SV03, PX4900SV02, HC0565PC03, TP0575UV02, GH1077UV01, 083055820116113, 083055824999515, 083055861021012, 83055862088813, 083055800046112, 81111DBS01, IL1469IT05, IL1469IT16, A5992711, A5412010, A6197102, KW0000PP02, HCIT4175024UV, AMS3963, BON0164, ECT0265, HCIT4192962UV, HCIT4264206UW, PL3673RW01, PL1493RW02, PL2970RW01, PL1124RW01, PL1499RW03, PL3349PC01, PL1501RW03, PL2624RW01, PL3249RW02, PL3277RW01, PL1005RW01, 600062UV, PL3362RW01, PL0326RW01, PL2975RW02, PL1504RW02, PL2971RW01, RU9255ID01, RU6968ID01, RU8945ID01, RU7519ID01, RU1781RW01, RU2580ID01, RU4255ID01, RU0003ID01, RU0472ID03, RU9595ID02, RU9327ID02, RU1011ID05, RU8366ID01, RU7808ID02, RU1516ID03, RU1646ID01, RU1112ID02, 01977813, 01980259, RU3052ID01, RU1107RW01, RU1107RW01, RU7175ID01, RU1717ID04, RU9216ID02, RU3453RW03, 600101RW02, RU8002ID01, RU1544ID01, RU1516ID03, RU9940RW01, RU5388ID01, RU0994ID03, RU3794ID01, RU9226ID01, RU9697ID01, 1509505, RU0968ID04, RU0120ID01, RU6164ID01, RU9546ID01, RU7468ID01, RU3998ID01, 600105ID03, SA2402PP02, SA2110PP05, SA1339PP03, ZA2541UV01, ZA2414PAC1, ZA2415PAC1, ZA2474UV01, ZA2267PAC1, ZA1058UV01, ZA2551PAC1, ZA1058UV01, ZA2412PAC2, ZA2472PAC2, ZA000010, ZA2412PAC2, MPX35219, DP760316, FP502804, 802030PC01, 228493WS01, 124774CS01, 482896SV03, 06051PAC04, HCIT4133734PACS, HCIT4148106PEE, Others Tracked under Service Cloud product ID. 2. Model Number: 2088026-723 (Centricity Universal Viewer with PACS-IW foundation software versions 6.0 through 6.0 SP7.1); UDI-DI: 840682103800; System ID: NYHOSPI16204605, NYHOSPI16204606, NYHOSPI16204607, NYHOSPI16204608, NYHOSPI16204609, NYHOSPI16204612, NYHOSPI16204613. 3. Model Number: 2102676-001 (Centricity Universal Viewer with PACS-IW foundation software versions 6.0 through 6.0 SP7.1); UDI-DI: 840682103800; System ID: PACS-IW-119. 4. Model Number: 2104867-044 (Centricity Universal Viewer with PACS-IW foundation software versions 6.0 through 6.0 SP7.1); UDI-DI: 840682103800; System ID: DDOC02339896.
AK, AZ, CA, CO, DE, FL, GA, IA, IN, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NJ, NM, NV, NY, OH, OK, OR, PA, SD, TX, UT, WA, WI, WV and WY. Austria, Brazil, Canada, France, Germany, Ghana, India, Ireland, Israel, Italy, Kuwait, Netherlands, Panama, Poland, Russia, Saudi Arabia, Slovenia, South Africa, Spain, Switzerland, Turkey, United Kingdom and Uruguay.
462 (279 US; 183 OUS)