FDA Enforcement Class II Terminated

Irix-C Forceps Inserter; Product Code/UDI: T066-0059/ M697T06600591

Recall: Z-2500-2020 · Reported July 15, 2020

Enforcement

Recall Number
Z-2500-2020
Event ID
85789
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
XTANT Medical
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
July 15, 2020
Initiation Date
May 12, 2020
Classification Date
July 4, 2020
Termination Date
March 19, 2021
Address
664 Cruiser Ln, Belgrade, MT, 59714-9719, United States

Description

Irix-C Forceps Inserter; Product Code/UDI: T066-0059/ M697T06600591

Reason

The strike plate fractured off from the forceps inserter devices with minimal impaction or force due to unsatisfactory weld.

Code Info

LOT: XS1920458A

Distribution

AZ, TX and WV.

Quantity

4 units