FDA Enforcement Class II Terminated

Threaded Offset Cup Impactor - Product Usage: To assist the surgeon in the implantation of Novation hip system components according to a conventional technique for total hip replacement. Specifically, these instruments are reusable, surgically invasive, and intended for transient use.

Recall: Z-2499-2019 · Reported September 18, 2019

Enforcement

Recall Number
Z-2499-2019
Event ID
83520
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Exactech, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
September 18, 2019
Initiation Date
July 24, 2019
Classification Date
September 10, 2019
Termination Date
May 31, 2023
Address
2320 NW 66th Ct, Gainesville, FL, 32653-1630, United States

Description

Threaded Offset Cup Impactor - Product Usage: To assist the surgeon in the implantation of Novation hip system components according to a conventional technique for total hip replacement. Specifically, these instruments are reusable, surgically invasive, and intended for transient use.

Reason

The threaded mechanism of the Threaded Offset Cup Impactor may be gritty and difficult to turn potentially due to entrapped blast media.

Code Info

Catalog Number 181-02-01, Lot Numbers 103071001, UDI Number 10885862520623. Expansion on 09/09/2019 to include Lot Number 8182001

Distribution

US distribution to AR, CA, FL, IL, IN, NY, SC, TX.

Quantity

22 devices