FDA Enforcement Class II Ongoing

MEDLINE Medical Procedure Kits labeled as: 1) CV DRAPING/BASIN, Medline Kit Number/SKU DYNJ908483A; 2) CV DRAPING/BASIN, Medline Kit Number/SKU DYNJ908483B; 3) CV DRAPING/BASIN, Medline Kit Number/SKU DYNJ908483C; 4) CV DRAPING/BASIN 2PART PACK, Medline Kit Number/SKU DYNJ89414; 5) CVG BASIN PACK SC, Medline Kit Number/SKU DYNJ38052G; 6) T/A BASIN SET, Medline Kit Number/SKU DYNJ47941F; 7) VALVE BASIN PACK SC, Medline Kit Number/SKU DYNJ38054G.

Recall: Z-2496-2026 · Reported June 24, 2026

Enforcement

Recall Number
Z-2496-2026
Event ID
98994
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Medline Industries, LP
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
June 24, 2026
Initiation Date
May 4, 2026
Classification Date
June 17, 2026
Address
3 Lakes Dr, Northfield, IL, 60093-2753, United States

Description

MEDLINE Medical Procedure Kits labeled as: 1) CV DRAPING/BASIN, Medline Kit Number/SKU DYNJ908483A; 2) CV DRAPING/BASIN, Medline Kit Number/SKU DYNJ908483B; 3) CV DRAPING/BASIN, Medline Kit Number/SKU DYNJ908483C; 4) CV DRAPING/BASIN 2PART PACK, Medline Kit Number/SKU DYNJ89414; 5) CVG BASIN PACK SC, Medline Kit Number/SKU DYNJ38052G; 6) T/A BASIN SET, Medline Kit Number/SKU DYNJ47941F; 7) VALVE BASIN PACK SC, Medline Kit Number/SKU DYNJ38054G.

Reason

During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.

Code Info

Medline Kit Number/SKU DYNJ908483A: UDI/DI 10195327191887, (each) 40195327191888, (case), Lot Number 22HMA818; Medline Kit Number/SKU DYNJ908483A: UDI/DI 10195327191887, (each) 40195327191888, (case), Lot Number 22GMG188; Medline Kit Number/SKU DYNJ908483B: UDI/DI 10195327287764, (each) 40195327287765, (case), Lot Number 23FMB021; Medline Kit Number/SKU DYNJ908483B: UDI/DI 10195327287764, (each) 40195327287765, (case), Lot Number 23CMB395; Medline Kit Number/SKU DYNJ908483B: UDI/DI 10195327287764, (each) 40195327287765, (case), Lot Number 23BMF105; Medline Kit Number/SKU DYNJ908483C: UDI/DI 10195327432119, (each) 40195327432110, (case), Lot Number 25BMG115; Medline Kit Number/SKU DYNJ908483C: UDI/DI 10195327432119, (each) 40195327432110, (case), Lot Number 24KMG981; Medline Kit Number/SKU DYNJ908483C: UDI/DI 10195327432119, (each) 40195327432110, (case), Lot Number 24KMG043; Medline Kit Number/SKU DYNJ908483C: UDI/DI 10195327432119, (each) 40195327432110, (case), Lot Number 24KME988; Medline Kit Number/SKU DYNJ908483C: UDI/DI 10195327432119, (each) 40195327432110, (case), Lot Number 24JMC883; Medline Kit Number/SKU DYNJ908483C: UDI/DI 10195327432119, (each) 40195327432110, (case), Lot Number 24HMG291; Medline Kit Number/SKU DYNJ908483C: UDI/DI 10195327432119, (each) 40195327432110, (case), Lot Number 24GMG957; Medline Kit Number/SKU DYNJ908483C: UDI/DI 10195327432119, (each) 40195327432110, (case), Lot Number 24GMC708; Medline Kit Number/SKU DYNJ908483C: UDI/DI 10195327432119, (each) 40195327432110, (case), Lot Number 24EMD960; Medline Kit Number/SKU DYNJ908483C: UDI/DI 10195327432119, (each) 40195327432110, (case), Lot Number 24EMD976; Medline Kit Number/SKU DYNJ908483C: UDI/DI 10195327432119, (each) 40195327432110, (case), Lot Number 24CMC199; Medline Kit Number/SKU DYNJ908483C: UDI/DI 10195327432119, (each) 40195327432110, (case), Lot Number 23LMH271; Medline Kit Number/SKU DYNJ908483C: UDI/DI 10195327432119, (each) 40195327432110, (case), Lot Number 23LMC861; Medline Kit Number/SKU DYNJ908483C: UDI/DI 10195327432119, (each) 40195327432110, (case), Lot Number 23JMI184; Medline Kit Number/SKU DYNJ908483C: UDI/DI 10195327432119, (each) 40195327432110, (case), Lot Number 23JMC583; Medline Kit Number/SKU DYNJ908483C: UDI/DI 10195327432119, (each) 40195327432110, (case), Lot Number 23IMH859; Medline Kit Number/SKU DYNJ908483C: UDI/DI 10195327432119, (each) 40195327432110, (case), Lot Number 23HME147; Medline Kit Number/SKU DYNJ908483C: UDI/DI 10195327432119, (each) 40195327432110, (case), Lot Number 23HMB823; Medline Kit Number/SKU DYNJ908483C: UDI/DI 10195327432119, (each) 40195327432110, (case), Lot Number 23GMD479; Medline Kit Number/SKU DYNJ908483C: UDI/DI 10195327432119, (each) 40195327432110, (case), Lot Number 23FMH321; Medline Kit Number/SKU DYNJ908483C: UDI/DI 10195327432119, (each) 40195327432110, (case), Lot Number 23FME722; Medline Kit Number/SKU DYNJ908483C: UDI/DI 10195327432119, (each) 40195327432110, (case), Lot Number 24EMC843; Medline Kit Number/SKU DYNJ89414: UDI/DI 10198459272127, (each) 40198459272128, (case), Lot Number 26CME546; Medline Kit Number/SKU DYNJ89414: UDI/DI 10198459272127, (each) 40198459272128, (case), Lot Number 26AMD910; Medline Kit Number/SKU DYNJ89414: UDI/DI 10198459272127, (each) 40198459272128, (case), Lot Number 25KMF065; Medline Kit Number/SKU DYNJ89414: UDI/DI 10198459272127, (each) 40198459272128, (case), Lot Number 25IMI592; Medline Kit Number/SKU DYNJ89414: UDI/DI 10198459272127, (each) 40198459272128, (case), Lot Number 25HMB017A; Medline Kit Number/SKU DYNJ89414: UDI/DI 10198459272127, (each) 40198459272128, (case), Lot Number 25HMB017; Medline Kit Number/SKU DYNJ89414: UDI/DI 10198459272127, (each) 40198459272128, (case), Lot Number 25GMK081; Medline Kit Number/SKU DYNJ89414: UDI/DI 10198459272127, (each) 40198459272128, (case), Lot Number 25DMI874; Medline Kit Number/SKU DYNJ89414: UDI/DI 10198459272127, (each) 40198459272128, (case), Lot Number 25BME130; Medline Kit Number/SKU DYNJ38052G: UDI/DI 10195327100131, (each) 40195327100132, (case), Lot Number 24CBE682; Medline Kit Number/SKU DYNJ38052G: UDI/DI 10195327100131, (each) 40195327100132, (case), Lot Number 23LBN771; Medline Kit Number/SKU DYNJ38052G: UDI/DI 10195327100131, (each) 40195327100132, (case), Lot Number 23LBJ194; Medline Kit Number/SKU DYNJ38052G: UDI/DI 10195327100131, (each) 40195327100132, (case), Lot Number 23KBO599; Medline Kit Number/SKU DYNJ38052G: UDI/DI 10195327100131, (each) 40195327100132, (case), Lot Number 23FBO284; Medline Kit Number/SKU DYNJ38052G: UDI/DI 10195327100131, (each) 40195327100132, (case), Lot Number 23EBM695; Medline Kit Number/SKU DYNJ38052G: UDI/DI 10195327100131, (each) 40195327100132, (case), Lot Number 22IBP839; Medline Kit Number/SKU DYNJ38052G: UDI/DI 10195327100131, (each) 40195327100132, (case), Lot Number 22GBU984; Medline Kit Number/SKU DYNJ38052G: UDI/DI 10195327100131, (each) 40195327100132, (case), Lot Number 22EBB509; Medline Kit Number/SKU DYNJ38052G: UDI/DI 10195327100131, (each) 40195327100132, (case), Lot Number 22DBE259; Medline Kit Number/SKU DYNJ38052G: UDI/DI 10195327100131, (each) 40195327100132, (case), Lot Number 22CBX946; Medline Kit Number/SKU DYNJ47941F: UDI/DI 10195327381370, (each) 40195327381371, (case), Lot Number 25CMI392; Medline Kit Number/SKU DYNJ47941F: UDI/DI 10195327381370, (each) 40195327381371, (case), Lot Number 24LMA643; Medline Kit Number/SKU DYNJ47941F: UDI/DI 10195327381370, (each) 40195327381371, (case), Lot Number 24EMD913; Medline Kit Number/SKU DYNJ47941F: UDI/DI 10195327381370, (each) 40195327381371, (case), Lot Number 23GMB073; Medline Kit Number/SKU DYNJ47941F: UDI/DI 10195327381370, (each) 40195327381371, (case), Lot Number 23FMF264; Medline Kit Number/SKU DYNJ47941F: UDI/DI 10195327381370, (each) 40195327381371, (case), Lot Number 23EMA833; Medline Kit Number/SKU DYNJ38054G: UDI/DI 10195327100148, (each) 40195327100149, (case), Lot Number 24CMA172; Medline Kit Number/SKU DYNJ38054G: UDI/DI 10195327100148, (each) 40195327100149, (case), Lot Number 23LMB006; Medline Kit Number/SKU DYNJ38054G: UDI/DI 10195327100148, (each) 40195327100149, (case), Lot Number 23KMF820; Medline Kit Number/SKU DYNJ38054G: UDI/DI 10195327100148, (each) 40195327100149, (case), Lot Number 23FMA119; Medline Kit Number/SKU DYNJ38054G: UDI/DI 10195327100148, (each) 40195327100149, (case), Lot Number 23AMG228; Medline Kit Number/SKU DYNJ38054G: UDI/DI 10195327100148, (each) 40195327100149, (case), Lot Number 22HMB147; Medline Kit Number/SKU DYNJ38054G: UDI/DI 10195327100148, (each) 40195327100149, (case), Lot Number 22BMH157.

Distribution

Worldwide distribution - US Nationwide and the countries of AE, BB, BM, BS, CA, CY, GU, NL, PA, QA, SA, VI.

Quantity

1432 units