FDA Enforcement Class II Terminated

Ingenuity Digital PET/CT System, Model 882446 with the following UDI's: (01)00884838059559(21)900004 (01)00884838059559(21)900005 (01)00884838059559(21)900006 (01)00884838059559(21)900007 (01)00884838059559(21)900008 (01)00884838059559(21)900009 (01)00884838059559(21)900010 (01)00884838059559(21)900011 (01)00884838059559(21)900012 (01)00884838059559(21)900013 (01)00884838059559(21)900014 (01)00884838059559(21)900015 (01)00884838059559(21)900016 (01)00884838059559(21)900017 (01)00884838059559(21)900018 (01)00884838059559(21)900019 (01)00884838059559(21)900020 (01)00884838059559(21)900021 (01)00884838059559(21)900022 (01)00884838059559(21)900023 (01)00884838059559(21)900024 (01)00884838059559(21)900025 (01)00884838059559(21)900026 (01)00884838059559(21)900027 (01)00884838059559(21)900028 (01)00884838059559(21)900029 (01)00884838059559(21)900030 (01)00884838059559(21)900031 (01)00884838059559(21)900032 (01)00884838059559(21)900033 (01)00884838059559(21)900034 (01)00884838059559(21)900035 Product Usage: The system is intended to image the whole body, heart, brain, lung, gastrointestinal, bone, lymphatic, and other major organs for a wide range of patient types, sizes, and extent of diseases. Both subsystems can also be operated as fully functional, independent diagnostic tools, including application of the CT scanner for diagnosis and for use in radiation therapy planning.

Recall: Z-2496-2018 · Reported July 25, 2018

Enforcement

Recall Number
Z-2496-2018
Event ID
80400
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Philips Medical Systems (Cleveland) Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
July 25, 2018
Initiation Date
May 21, 2018
Classification Date
July 19, 2018
Termination Date
April 10, 2020
Address
595 Miner Rd, Cleveland, OH, 44143-2131, United States

Description

Ingenuity Digital PET/CT System, Model 882446 with the following UDI's: (01)00884838059559(21)900004 (01)00884838059559(21)900005 (01)00884838059559(21)900006 (01)00884838059559(21)900007 (01)00884838059559(21)900008 (01)00884838059559(21)900009 (01)00884838059559(21)900010 (01)00884838059559(21)900011 (01)00884838059559(21)900012 (01)00884838059559(21)900013 (01)00884838059559(21)900014 (01)00884838059559(21)900015 (01)00884838059559(21)900016 (01)00884838059559(21)900017 (01)00884838059559(21)900018 (01)00884838059559(21)900019 (01)00884838059559(21)900020 (01)00884838059559(21)900021 (01)00884838059559(21)900022 (01)00884838059559(21)900023 (01)00884838059559(21)900024 (01)00884838059559(21)900025 (01)00884838059559(21)900026 (01)00884838059559(21)900027 (01)00884838059559(21)900028 (01)00884838059559(21)900029 (01)00884838059559(21)900030 (01)00884838059559(21)900031 (01)00884838059559(21)900032 (01)00884838059559(21)900033 (01)00884838059559(21)900034 (01)00884838059559(21)900035 Product Usage: The system is intended to image the whole body, heart, brain, lung, gastrointestinal, bone, lymphatic, and other major organs for a wide range of patient types, sizes, and extent of diseases. Both subsystems can also be operated as fully functional, independent diagnostic tools, including application of the CT scanner for diagnosis and for use in radiation therapy planning.

Reason

The block used to connect the patient pallet to the couch may be assembled incorrectly. A screw may be misaligned in the patient pallet.

Code Info

Serial Numbers: 900004 900005 900006 900007 900008 900009 900010 900011 900012 900013 900014 900015 900016 900017 900018 900019 900020 900021 900022 900023 900024 900025 900026 900027 900028 900029 900030 900031 900032 900033 900034 900035

Distribution

Worldwide Distribution - US Nationwide in the states of CA, GA, LA, MD, MI, NY, OH, TN, TX, VT, Washington DC & WI Internationally - Canada, Mexico, Belgium, France, Israel, Luxembourg, Netherlands, Singapore, Spain & Switzerland

Quantity

32 units