FDA Enforcement Class I Terminated

Breast Pack MRT, code 900-2658, contains: ( 1) BULB SYRINGE 60cc (2) GOWN SOFT SMS STD XLGE SET IN SLEEVE ( 1) CAUTERY PENCIL PUSH BOTTOM (2) SYRINGE 10cc W/0 NEEDLE LUER LOCK ( 1) NEEDLE and BLADE COUNTER 20c FOAM/MAG ( 1) UTILITY BOWL QUART 32oz ( 1) MAYO STAND COVER REINFORCED ( 1) TABLE COVER REINFORCED 50 X 90 ( 1) DRAPE BREAST/CHEST SMS ( 1) YANKAUER SUCTION TUBE W/0 VENT ( 1) TUBE SUCTION CONNECT Y ( 1) RULER 6 CLEAR (6) TOWEL CLOTH HUCK C (BLUE) (2) TOWEL ABSORBENT 15 X 20 (2) SKIN MARKER INK W/ LABEL ( 1) DENTURE CUP 8oz W/LID (2) LITE GLOVES Product Usage: EO sterilized surgical convenient packs packaged into polyester/polyethylene breather bag with Tyvek double vents.

Recall: Z-2491-2014 · Reported September 10, 2014

Enforcement

Recall Number
Z-2491-2014
Event ID
68536
Classification
Class I
Status
Terminated
Product Type
Devices
Firm
Customed, Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
September 10, 2014
Initiation Date
May 20, 2014
Classification Date
August 29, 2014
Termination Date
August 14, 2017
Address
Calle Igualdad Final #7, Fajardo, PR, 00738, United States

Description

Breast Pack MRT, code 900-2658, contains: ( 1) BULB SYRINGE 60cc (2) GOWN SOFT SMS STD XLGE SET IN SLEEVE ( 1) CAUTERY PENCIL PUSH BOTTOM (2) SYRINGE 10cc W/0 NEEDLE LUER LOCK ( 1) NEEDLE and BLADE COUNTER 20c FOAM/MAG ( 1) UTILITY BOWL QUART 32oz ( 1) MAYO STAND COVER REINFORCED ( 1) TABLE COVER REINFORCED 50 X 90 ( 1) DRAPE BREAST/CHEST SMS ( 1) YANKAUER SUCTION TUBE W/0 VENT ( 1) TUBE SUCTION CONNECT Y ( 1) RULER 6 CLEAR (6) TOWEL CLOTH HUCK C (BLUE) (2) TOWEL ABSORBENT 15 X 20 (2) SKIN MARKER INK W/ LABEL ( 1) DENTURE CUP 8oz W/LID (2) LITE GLOVES Product Usage: EO sterilized surgical convenient packs packaged into polyester/polyethylene breather bag with Tyvek double vents.

Reason

Customed has determined that there is the possibility that packaging integrity may be compromised on different catalog numbers (multiple lots). This could result in an injury to the patient due to product contamination or loss of sterility condition.

Code Info

Code 900-2658, 3 lots 112062217 112093595 112093720

Distribution

Worldwide Distribution - US (nationwide) in the state of Florida including Puerto Rico (USA); & USVI (St John) and the countries of Tortola, Trinidad, Dominican Republic, Costa Rica and Mexico.

Quantity

3 lots, 306 units, multiple units per lot